CLASS II Device recall · Reported 10 Jun 2020 · FDA Enforcement Report
Omnipod DASH Insulin Management System (mmol/L configuration) recalled
Insulet is recalling Omnipod DASH Insulin Management System (mmol/L configuration). The recall was initiated on 2 Apr 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2165-2020
- FDA event ID
- 85552
- Recalling firm
- Insulet, Acton, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 2 Apr 2020
- FDA classified
- 29 May 2020
- Reported
- 10 Jun 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Other Reasons
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 2 Apr 2020 | Recall initiated by company | |
| 29 May 2020 | Classified by FDA | +57 days |
| 10 Jun 2020 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.
Product as listed by the FDA
Omnipod DASH Insulin Management System (mmol/L configuration), Catalog Number PT000011 M/D: INT1D001MM - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Where it was distributed
The products were distributed to the following foreign countries: Italy, Netherlands, UK.
Questions about this recall
Why was it recalled?
After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.
Which lots are affected?
UDI: PT000011 10385082000146; Lot Numbers: L000200, L000201, L000203, L000209, L000210
Where was it sold?
The products were distributed to the following foreign countries: Italy, Netherlands, UK.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 85552This product is part of a recall event; the main page for the event is Omnipod DASH Insulin Management System (mg/dL configuration) recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Omnipod DASH Insulin Management System (mg/dL configuration) recalled | Insulet | Other | MA |
More recalls from Insulet
All 21| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Omnipod (Eros) Pod recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS I | Omnipod DASH Pod recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS I | Omnipod 5 Pods recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS I | Omnipod 5 Pods recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS II | The failure occurs if a user on the Omnipod 5 iOS recalled by Insulet | Insulet | Device Malfunction | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 467| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG Strip recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | ANNOUNCED | StatStrip Glucose Hospital Meter System recalled by Nova BiomedicalNova Biomedical StatStrip | Nova Biomedical | Device Malfunction | |
| Device | CLASS I | StatStrip Glucose Hospital Meter System recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS I | StatStrip Glucose and B- Ketone meter recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |