CLASS II Device recall · Reported 11 Feb 2015 · FDA Enforcement Report

Persona Tibial Articular Surface Inserter recalled by Zimmer

Zimmer is recalling Persona Tibial Articular Surface Inserter, distributed nationwide in the US. The recall was initiated on 25 Nov 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesItem No. 42-5299-001-00 Affected Lots: 56572679; 56573170; 56574318; 56575056; 56572918; 56573378; 56574538; 56575057; 56572975; 56573379; 56574540; 56575058; 56572978; 56574123; 56574618; 56575059; 56572981; 56574124; 56574619; 56575060; 56572982; 56574223; 56574620; 56575061; 56573071; 56574291; 56574629; 56575062; 56573115; 56574292; 56574630; 56575065; 56573116; 56574316; 56574631; 56575080; 56573154; 56574317; 56575018
Quantity recalled1,091

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1042-2015
FDA event ID
70105
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
25 Nov 2014
FDA classified
3 Feb 2015
Reported
11 Feb 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

25 Nov 2014Recall initiated by company
3 Feb 2015Classified by FDA+70 days
11 Feb 2015Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to contact with the fractured tip; Revision surgery to remove the fractured tip

Product as listed by the FDA

Zimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).

Where it was distributed

Worldwide distribution: US nationwide, including Puerto Rico, South Korea, Singapore, Dominican Republic, Canada, Japan, and Germany.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to contact with the fractured tip; Revision surgery to remove the fractured tip

Which lots are affected?

Item No. 42-5299-001-00 Affected Lots: 56572679; 56573170; 56574318; 56575056; 56572918; 56573378; 56574538; 56575057; 56572975; 56573379; 56574540; 56575058; 56572978; 56574123; 56574618; 56575059; 56572981; 56574124; 56574619; 56575060; 56572982; 56574223; 56574620; 56575061; 56573071; 56574291; 56574629; 56575062; 56573115; 56574292; 56574630; 56575065; 56573116; 56574316; 56574631; 56575080; 56573154; 56574317; 56575018

Where was it sold?

Worldwide distribution: US nationwide, including Puerto Rico, South Korea, Singapore, Dominican Republic, Canada, Japan, and Germany.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer

All 539

Other Hip & Knee Implants recalls

All 3,097