CLASS II Device recall · Reported 26 Oct 2016 · FDA Enforcement Report

Philips IntelliVue MX40 Patient Monitor recalled

Philips Electronics North America is recalling Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product, distributed nationwide in the US. The recall was initiated on 19 Sep 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSW Revisions: B.05.28, B.05.29, and B.05.32
Quantity recalled1824 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0292-2017
FDA event ID
75209
Recalling firm
Philips Electronics North America, Andover, MA
Classification
Class II
Status
Terminated
Recall initiated
19 Sep 2016
FDA classified
20 Oct 2016
Reported
26 Oct 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

19 Sep 2016Recall initiated by company
20 Oct 2016Classified by FDA+31 days
26 Oct 2016Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351 Worldwide Except USA Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 Exchange part: 453564262531 453564262551 453564262571 453564262591 453564615351 453564615371 453564467761 (Korea) 453564467781 (Korea) 453564451791 (Japan, Taiwan, Indonesia) 453564451811 (Japan, Taiwan, Indonesia) 453564467821 (Mexico) 453564467841 (China) 453564467861 (China) Worldwide except for the USA SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

Where it was distributed

Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.

Nationwide

Questions about this recall

Why was it recalled?

Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX

Which lots are affected?

SW Revisions: B.05.28, B.05.29, and B.05.32

Where was it sold?

Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.