CLASS II Device recall · Reported 6 Mar 2013 · FDA Enforcement Report

Philips Medical Systems, Extended Brilliance Workspace recalled

Philips Medical Systems (Cleveland) is recalling Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, distributed nationwide in the US. The recall was initiated on 24 Oct 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesThe EBW Serial Numbers subject to software updates are as follows: 8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127, 10331, 10348, 10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725, 11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957, and 11979.
Quantity recalled39 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0881-2013
FDA event ID
63822
Recalling firm
Philips Medical Systems (Cleveland), Cleveland, OH
Classification
Class II
Status
Terminated
Recall initiated
24 Oct 2012
FDA classified
26 Feb 2013
Reported
6 Mar 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

24 Oct 2012Recall initiated by company
26 Feb 2013Classified by FDA+125 days
6 Mar 2013Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

Product as listed by the FDA

Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System.

Where it was distributed

Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..

Nationwide AlaskaArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyMassachusettsMarylandMichiganMinnesotaMissouri

Questions about this recall

Is Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners recalled?

Yes. Philips Medical Systems (Cleveland) recalled Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners on 24 Oct 2012. The recall is Class II and its status is Terminated. Recall number Z-0881-2013.

Why was it recalled?

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

Which lots are affected?

The EBW Serial Numbers subject to software updates are as follows: 8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127, 10331, 10348, 10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725, 11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957, and 11979.

Where was it sold?

Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.