CLASS II ONGOING Device recall · Reported 29 Jul 2026 · FDA Enforcement Report

Pipeline Flex Embolization Device with Shield Technology recalled

Micro Therapeutics is recalling Pipeline Flex Embolization Device with Shield Technology 027, distributed nationwide in the US. The recall was initiated on 8 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 00763000448806/ Lot: D062612
Quantity recalled11 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2773-2026
FDA event ID
99326
Recalling firm
Micro Therapeutics, Irvine, CA
Classification
Class II
Status
Ongoing
Recall initiated
8 Jun 2026
FDA classified
22 Jul 2026
Reported
29 Jul 2026
Type
Voluntary: Firm initiated
Customers notified by
Visit
Reason category
Labeling Errors

Timeline

8 Jun 2026Recall initiated by company
22 Jul 2026Classified by FDA+44 days
29 Jul 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.

Product as listed by the FDA

Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12. Intracranial aneurysm flow diverter

Where it was distributed

US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.

Nationwide ConnecticutDelawareIndianaMassachusettsNorth CarolinaNew JerseyNew MexicoNew YorkOhioPennsylvaniaTexasVirginiaWashington

Questions about this recall

Why was it recalled?

Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.

Which lots are affected?

UDI: 00763000448806/ Lot: D062612

Where was it sold?

US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99326

This product is part of a recall event; the main page for the event is Pipeline Flex Embolization Device with Shield Technology recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIPipeline Flex Embolization Device with Shield Technology recalledMicro TherapeuticsLabeling ErrorsNationwide

More recalls from Micro Therapeutics

All 9

Other Other Medical Devices recalls

All 13,857