CLASS II Device recall · Reported 22 Jan 2014 · FDA Enforcement Report

pO2 D999 membrane units recalled by Radiometer America

Radiometer America is recalling pO2 D999 membrane units, distributed nationwide in the US. The recall was initiated on 10 Jun 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart #: 942-042, Lot #: R0971
Quantity recalled209 boxes/6 membranes

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0713-2014
FDA event ID
67044
Recalling firm
Radiometer America, Westlake, OH
Classification
Class II
Status
Terminated
Recall initiated
10 Jun 2013
FDA classified
10 Jan 2014
Reported
22 Jan 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

10 Jun 2013Recall initiated by company
10 Jan 2014Classified by FDA+214 days
22 Jan 2014Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.

Product as listed by the FDA

pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).

Where it was distributed

Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.

Nationwide ColoradoConnecticutGeorgiaMassachusettsMissouriMississippi

Questions about this recall

Why was it recalled?

RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.

Which lots are affected?

Part #: 942-042, Lot #: R0971

Where was it sold?

Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.