CLASS III Device recall · Reported 5 Jul 2017 · FDA Enforcement Report

Pointe Scientific CRP(HS) Wide Range Control Set recalled

Medtest Holdings is recalling Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range, distributed in 9 states. The recall was initiated on 21 Jul 2009 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesModel #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011.
Quantity recalled323 vials

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2631-2017
FDA event ID
77451
Recalling firm
Medtest Holdings, Canton, MI
Classification
Class III
Status
Terminated
Recall initiated
21 Jul 2009
FDA classified
26 Jun 2017
Reported
5 Jul 2017
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Labeling Errors

Timeline

21 Jul 2009Recall initiated by company
26 Jun 2017Classified by FDA+2897 days
5 Jul 2017Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.

Product as listed by the FDA

Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only

Where it was distributed

Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE

CaliforniaFloridaHawaiiMarylandMississippiNew JerseyNevadaOklahomaPennsylvania

Questions about this recall

Is Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range recalled?

Yes. Medtest Holdings recalled Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range on 21 Jul 2009. The recall is Class III and its status is Terminated. Recall number Z-2631-2017.

Why was it recalled?

Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.

Which lots are affected?

Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011.

Where was it sold?

Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.