CLASS II Device recall · Reported 9 Nov 2016 · FDA Enforcement Report
Product Number 3102071 Ceiling Bracket, Arch 71 White recalled
Hill-Rom is recalling Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail, distributed in Michigan, North Carolina, Pennsylvania, Washington. The recall was initiated on 9 Aug 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0326-2017
- FDA event ID
- 75197
- Recalling firm
- Hill-Rom, Batesville, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 Aug 2016
- FDA classified
- 31 Oct 2016
- Reported
- 9 Nov 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 9 Aug 2016 | Recall initiated by company | |
| 31 Oct 2016 | Classified by FDA | +83 days |
| 9 Nov 2016 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
Product as listed by the FDA
Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.
Where it was distributed
WA NC OH PA MI Australia Sweden
Questions about this recall
Why was it recalled?
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
Which lots are affected?
manufacturing date between Dec 01, 2015- Apr 14, 2016
Where was it sold?
WA NC OH PA MI Australia Sweden
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hill-Rom
All 37| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Traverse Rail Carrier E-System recalled by Hill-Rom | Hill-Rom | Other | Nationwide |
| Device | CLASS II | Liko Mobile Lifts: Primarily intended for use recalled by Hill-Rom | Hill-Rom | Other | Nationwide |
| Device | CLASS II | Liko Mobile Lifts: Primarily intended for use recalled by Hill-Rom | Hill-Rom | Other | Nationwide |
| Device | CLASS II | These are communication systems, labeled as recalled by Hill-Rom | Hill-Rom | Device Malfunction | Nationwide |
| Device | CLASS II | OctoStretch with Stretch Leveller recalled by Hill-Rom | Hill-Rom | Other | Nationwide |
Other Other Medical Devices recalls
All 13,929| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rapicide PA Part A High-Level Disinfectant recalled by Steris | Steris | Other | Nationwide |
| Device | CLASS II | NeuroStar Advanced Therapy System recalled by Neuronetics | Neuronetics | Other | Nationwide |
| Device | CLASS II | TS Misc. Instruments Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |