CLASS II Device recall · Reported 11 Jan 2017 · FDA Enforcement Report

Profemur(r) R Rasp Handle recalled by MicroPort Orthopedics

MicroPort Orthopedics is recalling Profemur(r) R Rasp Handle. The recall was initiated on 17 Nov 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number US100122
Quantity recalled20 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0943-2017
FDA event ID
75888
Recalling firm
MicroPort Orthopedics, Arlington, TN
Classification
Class II
Status
Terminated
Recall initiated
17 Nov 2016
FDA classified
3 Jan 2017
Reported
11 Jan 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Nov 2016Recall initiated by company
3 Jan 2017Classified by FDA+47 days
11 Jan 2017Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.

Product as listed by the FDA

PROFEMUR(R) R RASP HANDLE, REF PPW38078, 1 EACH, NON-STERILE, MicroPort Orthopedics, Arlington, TN Instruments support the implantation of implants. Broach handles mate to the broach to allow for preparation of the femoral canal for the stem component to be implanted. This

Where it was distributed

Worldwide Distribution to Spain, China, and Italy

Questions about this recall

Why was it recalled?

The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.

Which lots are affected?

Lot Number US100122

Where was it sold?

Worldwide Distribution to Spain, China, and Italy

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from MicroPort Orthopedics

All 31

Other Hip & Knee Implants recalls

All 3,111