CLASS II Device recall · Reported 18 Mar 2020 · FDA Enforcement Report

Qiagen Rotor Gene Q recalled by Qiagen Sciences

Qiagen Sciences is recalling Qiagen Rotor Gene Q, distributed nationwide in the US. The recall was initiated on 28 Jan 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware version 2.3.4 used in the following: Product No. / Description 9001680 Rotor-Gene Q 2Plex no PC 9001680R Rotor-Gene Q 2Plex no PC refurbished 9001690 Rotor-Gene Q 2Plex HRM no PC 9001690R Rotor-Gene Q 2Plex HRM no PC refurbished 9001700 Rotor-Gene Q 5Plex no PC 9001700R Rotor-Gene Q 5Plex no PC refurbished 9001710 Rotor-Gene Q 5Plex HRM no PC 9001710R Rotor-Gene Q 5Plex HRM no PC refurbished 9001720 Rotor-Gene Q 6Plex no PC 9001720R Rotor-Gene Q 6Plex no PC refurbished 9001960 Rotor-Gene AusDx (Outback Red) 9002000 Rotor-Gene Q MDx 2Plex 9002001 Rotor-Gene Q MDx 2Plex no PC 9002010 Rotor-Gene Q MDx 2Plex HRM 9002011 #Rotor-Gene Q MDx 2Plex HRM no PC 9002020 Rotor-Gene Q MDx 5Plex 9002021 Rotor-Gene Q MDx 5Plex no PC 9002029 Rotor-Gene Q MDx 5plex, PrioPLUS 9002030 Rotor-Gene Q MDx 5Plex HRM 9002030R Rotor-Gene Q MDx 5Plex HRM refurbished 9002031 Rotor-Gene Q MDx 5Plex HRM no PC 9002040 Rotor-Gene Q MDx 6Plex 9002041 Rotor-Gene Q MDx 6Plex no PC90275010 9002204 Rotor-Gene Q DBV 9023241 Rotor-Gene Q Software 9002370 5 plex MDx HRM with laptop (canada) This issue can affect the following QIAGEN IVD kits: Catalogue number Product Name 4514363 artus BK Virus QS-RGQ Kit (24) 4514263 artus BK Virus RG PCR Kit (24) 4514265 artus BK Virus RG PCR Kit (96) 4503363 artus CMV QS-RGQ Kit (24) 4503263 artus CMV RG PCR Kit (24) 4503265 artus CMV RG PCR Kit (96) 4501363 artus EBV QS-RGQ Kit (24) 4501263 artus EBV RG PCR Kit (24) 4501265 artus EBV RG PCR Kit (96) 4506366 artus HBV QS-RGQ (72) 4506356 artus HBV QS-RGQ Kit (24) 4506363 artus HBV QS-RGQ Kit (24) 4506263 artus HBV RG PCR Kit (24) 4506265 artus HBV RG PCR Kit (96) 4518356 artus HCV QS-RGQ Kit (24) 4518363 artus HCV QS-RGQ Kit (24) 4518366 artus HCV QS-RGQ Kit (72) 4538366 artus HCV QS-RGQ Kit V2 4518253 artus HCV RG RT-PCR Kit (24) 4518263 artus HCV RG RT-PCR Kit (24) 4518265 artus HCV RG RT-PCR Kit (96) 4513363 artus HI Virus-1 QS-RGQ Kit (24) 4513366 artus HI Virus-1 QS-RGQ Kit (72) 4513253 artus HI Virus-1 RG RT-PCR Kit (24) 4513263 artus HI Virus-1 RG RT-PCR Kit (24) 4513265 artus HI Virus-1 RG RT-PCR Kit (96) 4502363 artus VZV QS-RGQ Kit (24) 4502263 artus VZV RG PCR Kit (24) 4502265 artus VZV RG PCR Kit (96) 4532265 artus JCV RG PCR Kit (96) 4555263 artus M.tuberculosis RG PCR Kit (24) 4555265 artus M.tuberculosis RG PCR Kit (96) 4504263 artus Parvo B19 RG PCR Kit (24) 4511263 artus SARS RG RT-PCR Kit (24)
Quantity recalled11418 instruments

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1476-2020
FDA event ID
85004
Recalling firm
Qiagen Sciences, Germantown, MD
Classification
Class II
Status
Terminated
Recall initiated
28 Jan 2020
FDA classified
12 Mar 2020
Reported
18 Mar 2020
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

28 Jan 2020Recall initiated by company
12 Mar 2020Classified by FDA+44 days
18 Mar 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time detection of PCR assays.

Where it was distributed

Worldwide distribution - US Nationwide distribution and countries Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cameroon, Canada, Chile, China, Colombia, Congo, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Laos, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macao, Macedonia, the former Yugoslav Republic of, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mayotte, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestinian Territory Occupied, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab, Republic of Taiwan, Tajikistan, Thailand, Timor-Leste (else known as East Timor), Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic of, Viet Nam, Yemen.

Nationwide Georgia

Questions about this recall

Why was it recalled?

When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment

Which lots are affected?

Software version 2.3.4 used in the following: Product No. / Description 9001680 Rotor-Gene Q 2Plex no PC 9001680R Rotor-Gene Q 2Plex no PC refurbished 9001690 Rotor-Gene Q 2Plex HRM no PC 9001690R Rotor-Gene Q 2Plex HRM no PC refurbished 9001700 Rotor-Gene Q 5Plex no PC 9001700R Rotor-Gene Q 5Plex no PC refurbished 9001710 Rotor-Gene Q 5Plex HRM no PC 9001710R Rotor-Gene Q 5Plex HRM no PC refurbished 9001720 Rotor-Gene Q 6Plex no PC 9001720R Rotor-Gene Q 6Plex no PC refurbished 9001960 Rotor-Gene AusDx (Outback Red) 9002000 Rotor-Gene Q MDx 2Plex 9002001 Rotor-Gene Q MDx 2Plex no PC 9002010 Rotor-Gene Q MDx 2Plex HRM 9002011 #Rotor-Gene Q MDx 2Plex HRM no PC 9002020 Rotor-Gene Q MDx 5Plex 9002021 Rotor-Gene Q MDx 5Plex no PC 9002029 Rotor-Gene Q MDx 5plex, PrioPLUS 9002030 Rotor-Gene Q MDx 5Plex HRM 9002030R Rotor-Gene Q MDx 5Plex HRM refurbished 9002031 Rotor-Gene Q MDx 5Plex HRM no PC 9002040 Rotor-Gene Q MDx 6Plex 9002041 Rotor-Gene Q MDx 6Plex no PC90275010 9002204 Rotor-Gene Q DBV 9023241 Rotor-Gene Q Software 9002370 5 plex MDx HRM with laptop (canada) This issue can affect the following QIAGEN IVD kits: Catalogue number Product Name 4514363 artus BK Virus QS-RGQ Kit (24) 4514263 artus BK Virus RG PCR Kit (24) 4514265 artus BK Virus RG PCR Kit (96) 4503363 artus CMV QS-RGQ Kit (24) 4503263 artus CMV RG PCR Kit (24) 4503265 artus CMV RG PCR Kit (96) 4501363 artus EBV QS-RGQ Kit (24) 4501263 artus EBV RG PCR Kit (24) 4501265 artus EBV RG PCR Kit (96) 4506366 artus HBV QS-RGQ (72) 4506356 artus HBV QS-RGQ Kit (24) 4506363 artus HBV QS-RGQ Kit (24) 4506263 artus HBV RG PCR Kit (24) 4506265 artus HBV RG PCR Kit (96) 4518356 artus HCV QS-RGQ Kit (24) 4518363 artus HCV QS-RGQ Kit (24) 4518366 artus HCV QS-RGQ Kit (72) 4538366 artus HCV QS-RGQ Kit V2 4518253 artus HCV RG RT-PCR Kit (24) 4518263 artus HCV RG RT-PCR Kit (24) 4518265 artus HCV RG RT-PCR Kit (96) 4513363 artus HI Virus-1 QS-RGQ Kit (24) 4513366 artus HI Virus-1 QS-RGQ Kit (72) 4513253 artus HI Virus-1 RG RT-PCR Kit (24) 4513263 artus HI Virus-1 RG RT-PCR Kit (24) 4513265 artus HI Virus-1 RG RT-PCR Kit (96) 4502363 artus VZV QS-RGQ Kit (24) 4502263 artus VZV RG PCR Kit (24) 4502265 artus VZV RG PCR Kit (96) 4532265 artus JCV RG PCR Kit (96) 4555263 artus M.tuberculosis RG PCR Kit (24) 4555265 artus M.tuberculosis RG PCR Kit (96) 4504263 artus Parvo B19 RG PCR Kit (24) 4511263 artus SARS RG RT-PCR Kit (24)

Where was it sold?

Worldwide distribution - US Nationwide distribution and countries Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cameroon, Canada, Chile, China, Colombia, Congo, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Laos, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macao, Macedonia, the former Yugoslav Republic of, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mayotte, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestinian Territory Occupied, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab, Republic of Taiwan, Tajikistan, Thailand, Timor-Leste (else known as East Timor), Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic of, Viet Nam, Yemen.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Qiagen Sciences

All 23

Other Test Kits & Diagnostics recalls

All 2,675
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IINextseq 550Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IINovaSeq 6000Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IILiaison Plex Gram-Negative Blood Culture Assay recalled by LuminexLuminexPotency & AssayNationwide
DeviceCLASS IIMiSeq Dx Instrument recalled by IlluminaIlluminaOtherNationwide
DeviceCLASS IITechno-Path Multichem AE Unassayed Level recalledTechno-path ManufacturingPotency & AssayNationwide