CLASS III Device recall · Reported 2 Jan 2013 · FDA Enforcement Report
Qualicheck 1+, Aqueous Quality Control, Level 2, S7140 recalled
Radiometer America is recalling Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, distributed in Kentucky, North Carolina, New York. The recall was initiated on 6 Jun 2006 and the FDA has classified it as Class III.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0608-2013
- FDA event ID
- 37530
- Recalling firm
- Radiometer America, Westlake, OH
- Classification
- Class III
- Status
- Terminated
- Recall initiated
- 6 Jun 2006
- FDA classified
- 21 Dec 2012
- Reported
- 2 Jan 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Test Kits & Diagnostics
Timeline
| 6 Jun 2006 | Recall initiated by company | |
| 21 Dec 2012 | Classified by FDA | +2390 days |
| 2 Jan 2013 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Product as listed by the FDA
Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.
Where it was distributed
Distributed to KY, NC, and NY.
Questions about this recall
Why was it recalled?
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Which lots are affected?
Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.
Where was it sold?
Distributed to KY, NC, and NY.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class III mean?
Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Radiometer America
All 24| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | ABL90 FLEX Analyzer recalled by Radiometer America | Radiometer America | Device Malfunction | 36 states |
| Device | CLASS II | ABL800 FLEX with Crea recalled by Radiometer America | Radiometer America | Other | 31 states |
| Device | CLASS II | ABL800 analyzer recalled by Radiometer America | Radiometer America | Device Malfunction | Nationwide |
| Device | CLASS II | ABL90 FLEX analyzers recalled by Radiometer America | Radiometer America | Device Malfunction | CA |
| Device | CLASS II | ABL800 analyzer with Flexq module recalled by Radiometer America | Radiometer America | Other | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,661| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled | Biomerieux | Other | MO |
| Device | CLASS II | DASH Rapid PCR Instrument recalled by Nuclein | Nuclein | Device Malfunction | Nationwide |