CLASS II ONGOING Device recall · Reported 6 Nov 2024 · FDA Enforcement Report
Quest Spectrum 3 Fluorescent Imaging System recalled
Olympus Corporation of the Americas is recalling Quest Spectrum 3 Fluorescent Imaging System, distributed in Ohio. The recall was initiated on 24 Sep 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0232-2025
- FDA event ID
- 95423
- Recalling firm
- Olympus Corporation of the Americas, Center Valley, PA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 24 Sep 2024
- FDA classified
- 25 Oct 2024
- Reported
- 6 Nov 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 24 Sep 2024 | Recall initiated by company | |
| 25 Oct 2024 | Classified by FDA | +31 days |
| 6 Nov 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.
Where it was distributed
OH
Questions about this recall
Why was it recalled?
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
Which lots are affected?
UDI-DI: 08719324487047 Lot Code: Model: ART-SYS-0029, Serial: ART-SYS-2022-F04, ART-SYS-2022-A06, ART-SYS-2022-F05; Model: ART-SYS-0030; No distribution in the USA
Where was it sold?
OH
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Olympus Corporation of the Americas
All 415| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Powerseal (in diameter) (long), Curved Jaw Sealer & Divider recalled | Olympus Corporation of the Americas | Labeling Errors | Nationwide |
| Device | CLASS II | Olympus 4K UHD LCD Monitor recalled by Olympus Corporation of the | Olympus Corporation of the Americas | Other | Nationwide |
| Device | CLASS II | Olympus High Flow Insufflation Unit - UHI recalled | Olympus Corporation of the Americas | Other | Nationwide |
| Device | CLASS II | Olympus Toomey Evacuator 50cc with Adaptor recalled | Olympus Corporation of the Americas | Sterility | Nationwide |
| Device | CLASS II | Olympus Glass Body for 194 Ellik Evacuator recalled | Olympus Corporation of the Americas | Sterility | Nationwide |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederl | Potency & Assay |