CLASS II Device recall · Reported 31 Dec 2014 · FDA Enforcement Report

Replacement LED PCB Kit recalled by Natus Medical Incorporated

Natus Medical is recalling Replacement LED PCB Kit. The recall was initiated on 19 Nov 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)
Quantity recalled1907 total (1413 in US and 494 international)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0847-2015
FDA event ID
69820
Recalling firm
Natus Medical, Seattle, WA
Classification
Class II
Status
Terminated
Recall initiated
19 Nov 2014
FDA classified
22 Dec 2014
Reported
31 Dec 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

19 Nov 2014Recall initiated by company
22 Dec 2014Classified by FDA+33 days
31 Dec 2014Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Product as listed by the FDA

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

Where it was distributed

worldwide

Questions about this recall

Why was it recalled?

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Which lots are affected?

All distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)

Where was it sold?

worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Natus Medical

All 17

Other Other Medical Devices recalls

All 13,908