CLASS II ONGOING Device recall · Reported 18 Dec 2024 · FDA Enforcement Report

Rover Mobile X-ray System, # MXU-RV35 recalled by Micro-X

Micro-X is recalling Rover Mobile X-ray System, # MXU-RV35, distributed nationwide in the US. The recall was initiated on 31 May 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
Quantity recalled5

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0664-2025
FDA event ID
95603
Recalling firm
Micro-X, Tonsley
Classification
Class II
Status
Ongoing
Recall initiated
31 May 2023
FDA classified
12 Dec 2024
Reported
18 Dec 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

31 May 2023Recall initiated by company
12 Dec 2024Classified by FDA+561 days
18 Dec 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Product as listed by the FDA

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Where it was distributed

Worldwide and US Nationwide Distribution: NJ, PR, CA

Nationwide CaliforniaNew JerseyPuerto Rico

Questions about this recall

Why was it recalled?

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Which lots are affected?

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Where was it sold?

Worldwide and US Nationwide Distribution: NJ, PR, CA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Micro-X

All 10

Other Imaging Equipment recalls