CLASS II Device recall · Reported 19 Sep 2012 · FDA Enforcement Report
Safety monitor included with System 8000 Base The Sarns" recalled
Terumo Cardiovascular Systems is recalling Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is. The recall was initiated on 12 Aug 2009 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2348-2012
- FDA event ID
- 62321
- Recalling firm
- Terumo Cardiovascular Systems, Ann Arbor, MI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 12 Aug 2009
- FDA classified
- 11 Sep 2012
- Reported
- 19 Sep 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 12 Aug 2009 | Recall initiated by company | |
| 11 Sep 2012 | Classified by FDA | +1126 days |
| 19 Sep 2012 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.
Where it was distributed
Worldwide distribution: International only-country of: Terumo Dubai, UAE
Stores named: CVS
Questions about this recall
Why was it recalled?
Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav
Which lots are affected?
catalog number 16404 and serial numbers: 4357, 4381, 4386, 4387, 4389, 4391, and 4392.
Where was it sold?
Worldwide distribution: International only-country of: Terumo Dubai, UAE
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Terumo Cardiovascular Systems
All 222| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CDI OneView System BPM Probe recalled by Terumo Cardiovascular Systems | Terumo Cardiovascular Systems | Labeling Errors | Nationwide |
| Device | CLASS II | Terumo CDI OneView Monitoring System Hematocrit / Oxygen recalled | Terumo Cardiovascular Systems | Device Malfunction | 6 states |
| Device | CLASS II | Cardiovascular Procedure Kits that include High Crack Check recalled | Terumo Cardiovascular Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Capiox RX15 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
| Device | CLASS II | Capiox FX25 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |