CLASS II Device recall · Reported 27 Jun 2012 · FDA Enforcement Report
Sarns Flexible Arterial Cannula recalled
Terumo Cardiovascular Systems is recalling Sarns Flexible Arterial Cannula: With 3/8" Connector,suture Ring, 5" Long, distributed nationwide in the US. The recall was initiated on 9 May 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1837-2012
- FDA event ID
- 61742
- Recalling firm
- Terumo Cardiovascular Systems, Ann Arbor, MI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 May 2012
- FDA classified
- 20 Jun 2012
- Reported
- 27 Jun 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Heart & Vascular Devices
Timeline
| 9 May 2012 | Recall initiated by company | |
| 20 Jun 2012 | Classified by FDA | +42 days |
| 27 Jun 2012 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the parti
Product as listed by the FDA
SARNS FLEXIBLE ARTERIAL CANNULA: 8.0 MM (24 FR) WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
Where it was distributed
Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ , NM , NV , NY , OH , OK , OR , PA , RI , SC , SD , TN , TX , UT , VA , WA , WI, and WY and the countries of Australia, Belgium, Brazil, Canada, Chile, Germany, Honduras, Hong Kong, India, Japan, Mauritius, Nicaragua, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates (UAE), and Vietnam.
Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesota
Stores named: CVS
Questions about this recall
Is Sarns Flexible Arterial Cannula: With 3/8" Connector,suture Ring, 5" Long recalled?
Yes. Terumo Cardiovascular Systems recalled Sarns Flexible Arterial Cannula: With 3/8" Connector,suture Ring, 5" Long on 9 May 2012. The recall is Class II and its status is Terminated. Recall number Z-1837-2012.
Why was it recalled?
During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the parti
Which lots are affected?
Catalog number: 13030 and Lot code: 0553041, 0555967, 0557330, 0559885, 0562990, 0564951, 0567668, 0570456, 0571580, 0575365, 0580822, 0583186, 0585529, 0590858, 0593455, 0596125, 0600250, 0605504, 0609198, 0614183, 0620033, 0623325, 0626858, 0634096, 0636271, 0638011, 0640980, 0650900, 0653232, and 628985C.
Where was it sold?
Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ , NM , NV , NY , OH , OK , OR , PA , RI , SC , SD , TN , TX , UT , VA , WA , WI, and WY and the countries of Australia, Belgium, Brazil, Canada, Chile, Germany, Honduras, Hong Kong, India, Japan, Mauritius, Nicaragua, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates (UAE), and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Terumo Cardiovascular Systems
All 222| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CDI OneView System BPM Probe recalled by Terumo Cardiovascular Systems | Terumo Cardiovascular Systems | Labeling Errors | Nationwide |
| Device | CLASS II | Terumo CDI OneView Monitoring System Hematocrit / Oxygen recalled | Terumo Cardiovascular Systems | Device Malfunction | 6 states |
| Device | CLASS II | Cardiovascular Procedure Kits that include High Crack Check recalled | Terumo Cardiovascular Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Capiox RX15 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
| Device | CLASS II | Capiox FX25 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |