CLASS II Device recall · Reported 12 Jun 2013 · FDA Enforcement Report

SlingBar Wide recalled by Hill-Rom

Hill-Rom is recalling SlingBar Wide 670, distributed nationwide in the US. The recall was initiated on 30 Apr 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct # 3156012, 3156017
Quantity recalled412

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1475-2013
FDA event ID
64550
Recalling firm
Hill-Rom, Batesville, IN
Classification
Class II
Status
Terminated
Recall initiated
30 Apr 2013
FDA classified
5 Jun 2013
Reported
12 Jun 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

30 Apr 2013Recall initiated by company
5 Jun 2013Classified by FDA+36 days
12 Jun 2013Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.

Product as listed by the FDA

SlingBar Wide 670. Designed to meet the needs for lifting humans.

Where it was distributed

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.

Nationwide ArizonaCaliforniaColoradoFloridaIllinoisIndianaKansasLouisianaMassachusettsMichiganMinnesotaMissouriNorth CarolinaNew HampshireNew JerseyNevadaOhioOklahomaPennsylvaniaSouth Carolina

Questions about this recall

Why was it recalled?

Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.

Which lots are affected?

Product # 3156012, 3156017

Where was it sold?

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 64550

This product is part of a recall event; the main page for the event is SlingBar 350. Designed to meet the needs for lifting humans recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IISlingBar 350. Designed to meet the needs for lifting humans recalledHill-RomOtherNationwide
CLASS IISlingBar Standard recalled by Hill-RomHill-RomOtherNationwide
CLASS IISlingBar 360. Designed to meet the needs for lifting humans recalledHill-RomOtherNationwide

More recalls from Hill-Rom

All 37

Other Other Medical Devices recalls

All 13,929