CLASS II Device recall · Reported 26 Dec 2012 · FDA Enforcement Report

SOL SYS F 8 CALC LG Packaging recalled by DePuy Orthopaedics

DePuy Orthopaedics is recalling SOL SYS F 8 CALC LG Packaging: Product is packed within a poly protector, with protection, distributed nationwide in the US. The recall was initiated on 3 Dec 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Code: 157215136 and Lot codes: 169458, AB8F51000, AF1ED1000, AF1EDA000, AF1EG1000, AF1EGA000, AF1EGB000, AF7KH1000, AV7JX1000, AW8F11000, BR4J91000, BT9FM1000, BT9FMA000, BT9FMB000, C1PC61000, C35D51000, C35D5A000, C3WCX1000, C4YBF1000, C5FH51000, C5FH5A000, C77A71000, C77A7B, CG5B51000, CH1AE1000, CH1AEA000, CH1AEB000, CH1AF1000, D1TLE1000, D3RFF1000, D46LE1000, D52LY1000, D5FDJ1000, D74ER1000, DA8MC1000, DD6E51000, DG1N81000, DN5HY1000, DS7KB1000, E1RHH1000, E1RHHA000, E3SDE1000, E45LS1, E5LCW1, E65NEA, EB6DJ1000, EC3DJ1000, EF8HJ1000, EK8JR1000, ES6CV1000, EW2HF1000, EY2EG1000, FE9P31, FG2F21, FJ9L81, and FK9D41.
Quantity recalled1732 total all product codes

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0574-2013
FDA event ID
63545
Recalling firm
DePuy Orthopaedics, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
3 Dec 2012
FDA classified
20 Dec 2012
Reported
26 Dec 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

3 Dec 2012Recall initiated by company
20 Dec 2012Classified by FDA+17 days
26 Dec 2012Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

SOL SYS F 8 CALC 13.5MM LG Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Where it was distributed

Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichigan

Questions about this recall

Why was it recalled?

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Which lots are affected?

Product Code: 157215136 and Lot codes: 169458, AB8F51000, AF1ED1000, AF1EDA000, AF1EG1000, AF1EGA000, AF1EGB000, AF7KH1000, AV7JX1000, AW8F11000, BR4J91000, BT9FM1000, BT9FMA000, BT9FMB000, C1PC61000, C35D51000, C35D5A000, C3WCX1000, C4YBF1000, C5FH51000, C5FH5A000, C77A71000, C77A7B, CG5B51000, CH1AE1000, CH1AEA000, CH1AEB000, CH1AF1000, D1TLE1000, D3RFF1000, D46LE1000, D52LY1000, D5FDJ1000, D74ER1000, DA8MC1000, DD6E51000, DG1N81000, DN5HY1000, DS7KB1000, E1RHH1000, E1RHHA000, E3SDE1000, E45LS1, E5LCW1, E65NEA, EB6DJ1000, EC3DJ1000, EF8HJ1000, EK8JR1000, ES6CV1000, EW2HF1000, EY2EG1000, FE9P31, FG2F21, FJ9L81, and FK9D41.

Where was it sold?

Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

19 other products · FDA event 63545

This product is part of a recall event; the main page for the event is SOL SYS CALC 1.5/18.0 Packaging recalled by DePuy Orthopaedics.

ReportedClassRecallCompanyReasonWhere
CLASS IISOL SYS CALC 1.5/18.0 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS L CALC 2.25/15.0 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS R CALC 2.25/16.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS CALC 1.5/19.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS L CALC 2.25/19.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS R CALC 2.25/18 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS R CALC 2.25/21 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS R CALC 2.25/19.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS L CALC 2.25/22.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS F 9 CALC L LG Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS R CALC 2.25/15.0 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS CALC 1.5/16.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS CALC 1.5/15.0 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS L CALC 2.25/21 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS L CALC 2.25/18 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS R CALC 2.25/22.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS F 9 CALC L LG Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS L CALC 2.25/16.5 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide
CLASS IISOL SYS Calcar SZ12 Packaging recalled by DePuy OrthopaedicsDePuy OrthopaedicsDevice MalfunctionNationwide

More recalls from DePuy Orthopaedics

All 153

Other Hip & Knee Implants recalls

All 3,097