CLASS II Device recall · Reported 17 Aug 2016 · FDA Enforcement Report

Stryker Orthopaedics Modular Handle Triathlon Instruments recalled

Stryker Howmedica Osteonics is recalling Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, distributed nationwide in the US. The recall was initiated on 23 Jun 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog No. 6541-4-808 Lot Nos.: SB1K14, SB1K14A, SB1K14A1, SB1K14P, SB1K15, SB1K15A, SB1K15A1, SB1K15A2, SB1M25, SB1M25A, SB1M25A1, SB1M25A2, SB1M25A2X1, SB1N02, SB1N02A, SB1N02A1, SB1N02A2, SB1N02A3, SB1N02X1, SB1T05, SB1T05A, SB1T05A1, SB1T05AX1, SB1T05AX2, SB1T05KSB1V01, SB1V01K, SB1V01L, SB1V01X1, SB1V01X2, SB1V01X3, SB1W14, SB1W14K, SB1W14X1, SB2T19, SB2T19A, SB2T19D, SB2T19DX1, SB2T19DX2, SB2T19J, SB2T19K, SB2T19KX1, SB2T19KX2, SBT19X3, SB2V87, SB2V87A, SB2V87J, SB2V87JX1, SB2V87K, SB2V87X, SB2V87X1, SB2W23, SB2W23G, SB2W23X, SB2W23X1, SB3A20, SB3A20A, SB3A21, SB3A21D, SB3A21M, SB3A21X, SB3A22, SB3A22A1, SB3A22X1, SB3A55, SB3A55A, SB3A55A2, SB3A55AX1, SB3A55M, SB3A55X1, SB3H25, SB3H25D, SB3H25G, SB3H25X1, SB3H26, SB3H26D, SB3H26DX1, SB3H26X2, SB3K09, SB3K09X1, SB3K10, SB3K10L, SB3K10M, SB3K10MX1, SB3K10X1, SB3L19, SB3L19D, SB3L19K, SB3L19L, SB3L19P, SB3L19P1, SB3L20, SB3L20K, SB3L45, SB3L45A, SB3L45A1, SB3L45A2, SB3L45D, SB3L45DX1, SB3L45L, SB3L45T, SB3M17, SB3M17A, SB3M17K, SB3M17M, SB3M18, SB3M18A, SB3M18AX1, SB3M18D, SB3M18M, SB3M18R, SB3M19, SB3M19A, SB3M19A1, SB3M19S, SB3M19T, SB3M19W, SB3N46, SB3N46D, SB3N46P, SB3N46X1, SB4C12, SB4C12X1, SB4C37, SB4C37X1, SB4C38, SB4C38A, SB4C38D, SB4E18, SB4E181, SB4E181X1, SB4E18A, SB4E18D, SB4E18T, SB4H79, SB4H79A, SB4H79M, SB4L05, SB4L05A, SB4N17, SB4N17E, SB4N17M, SB4N17M1, SB4N18, SB4N18A, SB4N18A1, SB4N18D, SB4N18M, SB4N18T, SB4S01, SB4S01D, SB4S01J, SB4V07, SB4V07L, SB4V07M, SB4V07MM, SB5A02, SB5A02A, SB5A02A1, SB5A55
Quantity recalled12,469 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2377-2016
FDA event ID
74634
Recalling firm
Stryker Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Terminated
Recall initiated
23 Jun 2016
FDA classified
9 Aug 2016
Reported
17 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Sterility

Timeline

23 Jun 2016Recall initiated by company
9 Aug 2016Classified by FDA+47 days
17 Aug 2016Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.

Where it was distributed

Worldwide Distribution - US Nationwide

Nationwide

Questions about this recall

Is Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile recalled?

Yes. Stryker Howmedica Osteonics recalled Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile on 23 Jun 2016. The recall is Class II and its status is Terminated. Recall number Z-2377-2016.

Why was it recalled?

Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.

Which lots are affected?

Catalog No. 6541-4-808 Lot Nos.: SB1K14, SB1K14A, SB1K14A1, SB1K14P, SB1K15, SB1K15A, SB1K15A1, SB1K15A2, SB1M25, SB1M25A, SB1M25A1, SB1M25A2, SB1M25A2X1, SB1N02, SB1N02A, SB1N02A1, SB1N02A2, SB1N02A3, SB1N02X1, SB1T05, SB1T05A, SB1T05A1, SB1T05AX1, SB1T05AX2, SB1T05KSB1V01, SB1V01K, SB1V01L, SB1V01X1, SB1V01X2, SB1V01X3, SB1W14, SB1W14K, SB1W14X1, SB2T19, SB2T19A, SB2T19D, SB2T19DX1, SB2T19DX2, SB2T19J, SB2T19K, SB2T19KX1, SB2T19KX2, SBT19X3, SB2V87, SB2V87A, SB2V87J, SB2V87JX1, SB2V87K, SB2V87X, SB2V87X1, SB2W23, SB2W23G, SB2W23X, SB2W23X1, SB3A20, SB3A20A, SB3A21, SB3A21D, SB3A21M, SB3A21X, SB3A22, SB3A22A1, SB3A22X1, SB3A55, SB3A55A, SB3A55A2, SB3A55AX1, SB3A55M, SB3A55X1, SB3H25, SB3H25D, SB3H25G, SB3H25X1, SB3H26, SB3H26D, SB3H26DX1, SB3H26X2, SB3K09, SB3K09X1, SB3K10, SB3K10L, SB3K10M, SB3K10MX1, SB3K10X1, SB3L19, SB3L19D, SB3L19K, SB3L19L, SB3L19P, SB3L19P1, SB3L20, SB3L20K, SB3L45, SB3L45A, SB3L45A1, SB3L45A2, SB3L45D, SB3L45DX1, SB3L45L, SB3L45T, SB3M17, SB3M17A, SB3M17K, SB3M17M, SB3M18, SB3M18A, SB3M18AX1, SB3M18D, SB3M18M, SB3M18R, SB3M19, SB3M19A, SB3M19A1, SB3M19S, SB3M19T, SB3M19W, SB3N46, SB3N46D, SB3N46P, SB3N46X1, SB4C12, SB4C12X1, SB4C37, SB4C37X1, SB4C38, SB4C38A, SB4C38D, SB4E18, SB4E181, SB4E181X1, SB4E18A, SB4E18D, SB4E18T, SB4H79, SB4H79A, SB4H79M, SB4L05, SB4L05A, SB4N17, SB4N17E, SB4N17M, SB4N17M1, SB4N18, SB4N18A, SB4N18A1, SB4N18D, SB4N18M, SB4N18T, SB4S01, SB4S01D, SB4S01J, SB4V07, SB4V07L, SB4V07M, SB4V07MM, SB5A02, SB5A02A, SB5A02A1, SB5A55

Where was it sold?

Worldwide Distribution - US Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Howmedica Osteonics

All 86

Other Hip & Knee Implants recalls

All 3,097