CLASS II Device recall · Reported 29 Nov 2017 · FDA Enforcement Report
Syngo.plaza, Picture Archiving and Communications System(PACS) recalled
Siemens Medical Solutions USA is recalling Syngo.plaza, Picture Archiving and Communications System(PACS), distributed in Minnesota. The recall was initiated on 22 May 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0168-2018
- FDA event ID
- 77456
- Recalling firm
- Siemens Medical Solutions USA, Malvern, PA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 22 May 2017
- FDA classified
- 21 Nov 2017
- Reported
- 29 Nov 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 22 May 2017 | Recall initiated by company | |
| 21 Nov 2017 | Classified by FDA | +183 days |
| 29 Nov 2017 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Where it was distributed
Distribution to MN
Questions about this recall
Why was it recalled?
Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.
Which lots are affected?
non-medical software application Check Tool V1 .0 (L TA Incomplete Archive Check Tool V1 .0)
Where was it sold?
Distribution to MN
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Medical Solutions USA
All 796| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono biplane recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono floor recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis pheno recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | PA |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |