CLASS II Device recall · Reported 19 Jun 2019 · FDA Enforcement Report
Terumo Advanced Perfusion System 1 Electronic Patient Gas recalled
Terumo Cardiovascular Systems is recalling Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas, distributed nationwide in the US. The recall was initiated on 17 Oct 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1769-2019
- FDA event ID
- 82878
- Recalling firm
- Terumo Cardiovascular Systems, Ann Arbor, MI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 17 Oct 2018
- FDA classified
- 11 Jun 2019
- Reported
- 19 Jun 2019
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 17 Oct 2018 | Recall initiated by company | |
| 11 Jun 2019 | Classified by FDA | +237 days |
| 19 Jun 2019 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588. Used in cardio-pulmonary bypass.
Where it was distributed
Distribution US nationwide, including CA, CT, IL, IN, KY, MI, MN, MO, MT, NM, NY, OH, and WI. Foreign distribution to Canada, Mexico, and South Korea.
Nationwide CaliforniaConnecticutIllinoisIndianaKentuckyMichiganMinnesotaMissouriMontanaNew MexicoNew YorkOhioWisconsin
Questions about this recall
Why was it recalled?
An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.
Which lots are affected?
01338 01339 01340 01341 01342 01349 01350 01351 01352 01353 01354 01355 01356 01357 01358 01360 01361 01362 01363 01365 01366 01367 01369 01370 01371 01372 01373 01374 01375 01376 01377 01378 01379 Lot/Serial 01380 01381 01382 01383 01384 01385 01386 01387 01388 01389 01394 01395 01396 01397 01398 01405 01406 01407 01408 01410 01411 01413 01414 01415 01416 01418 01420 01421 01422 01423 01424 01425 01426 Lot/Serial 01427 01428 01429 01430 01431 01432 01434 01435 1718 2317 4417 9999 UDI 00886799000588
Where was it sold?
Distribution US nationwide, including CA, CT, IL, IN, KY, MI, MN, MO, MT, NM, NY, OH, and WI. Foreign distribution to Canada, Mexico, and South Korea.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Terumo Cardiovascular Systems
All 222| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CDI OneView System BPM Probe recalled by Terumo Cardiovascular Systems | Terumo Cardiovascular Systems | Labeling Errors | Nationwide |
| Device | CLASS II | Terumo CDI OneView Monitoring System Hematocrit / Oxygen recalled | Terumo Cardiovascular Systems | Device Malfunction | 6 states |
| Device | CLASS II | Cardiovascular Procedure Kits that include High Crack Check recalled | Terumo Cardiovascular Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Capiox RX15 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
| Device | CLASS II | Capiox FX25 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
Other Other Medical Devices recalls
All 13,884| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | Nationwide |
| Device | CLASS II | Palacos R pro recalled by Heraeus Medical GmbH (Dental | Heraeus Medical GmbH (Dental Division) | Packaging Defects | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065 recalled | Percussionaire | Other | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV-1C recalled by Percussionaire | Percussionaire | Other | Nationwide |