CLASS II Device recall · Reported 24 Oct 2018 · FDA Enforcement Report

Terumo¿ Advanced Perfusion System 1 Electronic Patient Gas System recalled

Terumo Cardiovascular Systems is recalling Terumo¿ Advanced Perfusion System 1 Electronic Patient Gas System, distributed nationwide in the US. The recall was initiated on 13 Sep 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll systems. UDI 00886799000588
Quantity recalled152

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0176-2019
FDA event ID
81070
Recalling firm
Terumo Cardiovascular Systems, Ann Arbor, MI
Classification
Class II
Status
Terminated
Recall initiated
13 Sep 2018
FDA classified
18 Oct 2018
Reported
24 Oct 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

13 Sep 2018Recall initiated by company
18 Oct 2018Classified by FDA+35 days
24 Oct 2018Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The service manual was not updated at the time of the release of software version 1.30.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Terumo¿ Advanced Perfusion System 1 Electronic Patient Gas System, Catalog#: 801188

Where it was distributed

US Nationwide in the states of: AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, MI, MN, MO, OH, OR, PA, SC, TN, TX, WA, WI, and Puerto Rico. The products were distributed to the following foreign countries: Antioquia, Australia, Bangkok, Bangladesh, Belgium, Canada, Chile, China, Columbia, Dominican Republic, Ecuador, El Salvador, France, Germany, Guatemala, India, Indonesia, Japan, Korea, Kuwait, Malaysia, Metro Manilla, Mexico, New Delhi, New Zealand, Panama, Philippines, Singapore, South Africa, Sri Lanka, Taiwan, Thailand, and Vietnam.

Nationwide AlabamaArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIllinoisIndianaKansasKentuckyMichiganMinnesotaMissouriOhioOregonPennsylvaniaPuerto RicoSouth CarolinaTennessee

Questions about this recall

Why was it recalled?

The service manual was not updated at the time of the release of software version 1.30.

Which lots are affected?

All systems. UDI 00886799000588

Where was it sold?

US Nationwide in the states of: AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, MI, MN, MO, OH, OR, PA, SC, TN, TX, WA, WI, and Puerto Rico. The products were distributed to the following foreign countries: Antioquia, Australia, Bangkok, Bangladesh, Belgium, Canada, Chile, China, Columbia, Dominican Republic, Ecuador, El Salvador, France, Germany, Guatemala, India, Indonesia, Japan, Korea, Kuwait, Malaysia, Metro Manilla, Mexico, New Delhi, New Zealand, Panama, Philippines, Singapore, South Africa, Sri Lanka, Taiwan, Thailand, and Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Terumo Cardiovascular Systems

All 222

Other Other Medical Devices recalls

All 13,884