CLASS II Device recall · Reported 1 Sep 2021 · FDA Enforcement Report

Terumo Advanced Perfusion System 1 Electronic Patient Gas System recalled

Terumo Cardiovascular Systems is recalling Terumo Advanced Perfusion System 1 Electronic Patient Gas System. The recall was initiated on 8 Jun 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesS/N: 01751 UDI: 00886799000588
Quantity recalled1 unit OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2350-2021
FDA event ID
88350
Recalling firm
Terumo Cardiovascular Systems, Ann Arbor, MI
Classification
Class II
Status
Terminated
Recall initiated
8 Jun 2021
FDA classified
26 Aug 2021
Reported
1 Sep 2021
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

8 Jun 2021Recall initiated by company
26 Aug 2021Classified by FDA+79 days
1 Sep 2021Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188

Where it was distributed

International distribution in the country of Panama.

Questions about this recall

Why was it recalled?

intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs

Which lots are affected?

S/N: 01751 UDI: 00886799000588

Where was it sold?

International distribution in the country of Panama.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.