CLASS II Device recall · Reported 11 May 2022 · FDA Enforcement Report
Velys Robotic-Assisted Solution Base Product No. recalled
DePuy Orthopaedics is recalling Velys Robotic-Assisted Solution Base Product No.: 451570100, distributed nationwide in the US. The recall was initiated on 11 Mar 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1029-2022
- FDA event ID
- 89959
- Recalling firm
- DePuy Orthopaedics, Warsaw, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 11 Mar 2022
- FDA classified
- 2 May 2022
- Reported
- 11 May 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 11 Mar 2022 | Recall initiated by company | |
| 2 May 2022 | Classified by FDA | +52 days |
| 11 May 2022 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
VELYS Robotic-Assisted Solution Base Product No.: 451570100
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.
Questions about this recall
Why was it recalled?
System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment
Which lots are affected?
Serial Numbers (US): BAS1111 BAS1145 BAS1154 BAS1069 BAS1070 BAS1125 BAS1144 BAS1174 BAS1175 BAS1192 BAS1194 BAS1081 BAS1082 BAS1052 BAS1165 BAS1136 BAS1071 BAS1142 BAS1073 BAS1189 BAS1022 BAS1190 BAS1079 BAS1054 BAS1061 BAS1173 BAS1204 BAS1141 BAS1076 BAS1181 BAS1176 BAS1184 BAS1119 BAS1137 BAS1009 BAS1140 BAS1121 BAS1143 BAS1127 BAS1060 BAS1004 BAS1196 BAS1120 BAS1068 BAS1152 BAS1188 BAS1078 BAS1195 BAS1191 BAS1062 BAS1180 BAS1118 BAS1197 BAS1139 BAS1066 BAS1126 BAS1093 BAS1122 BAS1134 BAS1203 BAS1132 BAS1185 BAS1182 BAS1129 BAS1063 BAS1183 BAS1133 BAS1212 UDI: 10603295519515
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from DePuy Orthopaedics
All 153| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Attune Affixium Cementless Fixed Bearing Knee recalled | DePuy Orthopaedics | Labeling Errors | Nationwide |
| Device | CLASS II | Attune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 recalled | DePuy Orthopaedics | Other | Nationwide |
| Device | CLASS II | Attune Revision Limb Preservation System (LPS) Tibial Insert XSM recalled | DePuy Orthopaedics | Other | Nationwide |
| Device | CLASS II | Attune Revision Limb Preservation System (LPS) Tibial Insert XSM recalled | DePuy Orthopaedics | Other | Nationwide |
| Device | CLASS II | Attune Revision Limb Preservation System (LPS) Tibial Insert SM recalled | DePuy Orthopaedics | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |