CLASS II ONGOING Device recall · Reported 14 Oct 2026 · FDA Enforcement Report

Verify Dual Species Self-Contained Biological Indicator recalled

Steris is recalling Verify Dual Species Self-Contained Biological Indicator, distributed nationwide in the US. The recall was initiated on 31 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 10724995023833 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027
Quantity recalled1185

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0072-2027
FDA event ID
99704
Recalling firm
Steris, Mentor, OH
Classification
Class II
Status
Ongoing
Recall initiated
31 Aug 2026
FDA classified
2 Oct 2026
Reported
14 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

31 Aug 2026Recall initiated by company
2 Oct 2026Classified by FDA+32 days
14 Oct 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Which lots are affected?

UDI: 10724995023833 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 99704

This product is part of a recall event; the main page for the event is Verify Biological Steam Test Pack recalled by Steris.