CLASS II Device recall · Reported 31 Oct 2018 · FDA Enforcement Report

VirtuoSaph Plus Endoscopic Vessel Harvesting System recalled

Terumo Cardiovascular Systems is recalling VirtuoSaph Plus Endoscopic Vessel Harvesting System, distributed nationwide in the US. The recall was initiated on 31 Aug 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Numbers: 81K
Quantity recalled135

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0227-2019
FDA event ID
81094
Recalling firm
Terumo Cardiovascular Systems, Elkton, MD
Classification
Class II
Status
Terminated
Recall initiated
31 Aug 2018
FDA classified
23 Oct 2018
Reported
31 Oct 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

31 Aug 2018Recall initiated by company
23 Oct 2018Classified by FDA+53 days
31 Oct 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has been removed from the list of compatible generators within the IFU.

Product as listed by the FDA

VirtuoSaph Plus Endoscopic Vessel Harvesting System, Catalog Number: VSP550EX Product Usage: The VittuoSaph¿ Plus Endoscopic Vessel Harvesting System VSP550EX is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surge1y for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requit¿ing blunt dissection of tissue including dissection of blood vessels and dissection of blood vessels of the extremities. Extremity procedures include tissue dissection and/or vessel harvesting along the saphenous vein for coronary artery bypass grafting and peripheral artery bypass grafting or radial artery for use in coronary artery bypass grafting vc

Where it was distributed

Worldwide - US Nationwide in the states of FL, IL, NY, OH, TX. The products were distributed to the following foreign country: Singapore.

Nationwide FloridaIllinoisNew YorkOhioTexas

Questions about this recall

Why was it recalled?

The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has been removed from the list of compatible generators within the IFU.

Which lots are affected?

Lot Numbers: 81K

Where was it sold?

Worldwide - US Nationwide in the states of FL, IL, NY, OH, TX. The products were distributed to the following foreign country: Singapore.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Terumo Cardiovascular Systems

All 222

Other Surgical Instruments recalls

All 3,883