CLASS III Device recall · Reported 28 Dec 2016 · FDA Enforcement Report

Vitros Crp Slides: Vitros¿ Chemistry Products Crp Slides recalled

Ortho-Clinical Diagnostics is recalling Vitros Crp Slides: Vitros¿ Chemistry Products Crp Slides (250 slides per sales unit), distributed nationwide in the US. The recall was initiated on 26 Oct 2016 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesAffected Generations (GENs): 3, 11, 12, 30 through 36
Quantity recalled124,063 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0864-2017
FDA event ID
75692
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class III
Status
Terminated
Recall initiated
26 Oct 2016
FDA classified
21 Dec 2016
Reported
28 Dec 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

26 Oct 2016Recall initiated by company
21 Dec 2016Classified by FDA+56 days
28 Dec 2016Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.

Product as listed by the FDA

VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Products CRP Slides (90 slides per sales unit), Catalog # 1926740

Where it was distributed

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.

Which lots are affected?

Affected Generations (GENs): 3, 11, 12, 30 through 36

Where was it sold?

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Other Medical Devices recalls

All 13,778