CLASS II ONGOING Device recall · Reported 28 Aug 2024 · FDA Enforcement Report
The Vitros XT Chemistry Products Alb-tp Slides is a single recalled
Ortho-Clinical Diagnostics is recalling The Vitros XT Chemistry Products Alb-tp Slides is a single device that contains both an, distributed nationwide in the US. The recall was initiated on 10 Jul 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2684-2024
- FDA event ID
- 94993
- Recalling firm
- Ortho-Clinical Diagnostics, Rochester, NY
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 10 Jul 2024
- FDA classified
- 16 Aug 2024
- Reported
- 28 Aug 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Potency & Assay
- Device type
- Test Kits & Diagnostics
Timeline
| 10 Jul 2024 | Recall initiated by company | |
| 16 Aug 2024 | Classified by FDA | +37 days |
| 28 Aug 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test side by side separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur, and test results are generated for each analyte independently of the other analyte. The ALB test is a multilayered, analytical element coated on a polyester support. For the albumin measurement, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. When the fluid penetrates the reagent layer, the bromocresol green (BCG) dye diffuses to the spreading layer and binds to albumin in the sample. This binding results in a shift in wavelength of the reflectance maximum of the free dye. The color complex that forms is measured by reflectance spectrophotometry. The amount of albumin-bound dye is proportional to the concentration of albumin in the sample. Model/Catalog Number: 6844293
Where it was distributed
Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden The Netherlands United Kingdom
Questions about this recall
Why was it recalled?
The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.
Which lots are affected?
UDI/GTIN# 10758750031771 All current and expired coatings
Where was it sold?
Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden The Netherlands United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Ortho-Clinical Diagnostics
All 311| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Chemistry Products LAC Slides recalled by Ortho-Clinical Diagnostics | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products LAC Slides recalled | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products Ca Slides recalled | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products CHE Slides (Japan) recalled | Ortho-Clinical Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Vitros Chemistry Products CHE Slides recalled | Ortho-Clinical Diagnostics | Labeling Errors | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,661| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled | Biomerieux | Other | MO |
| Device | CLASS II | DASH Rapid PCR Instrument recalled by Nuclein | Nuclein | Device Malfunction | Nationwide |