CLASS II ONGOING Device recall · Reported 27 Sep 2023 · FDA Enforcement Report

Vitros XT Chemistry Products Alkp Slides recalled

Ortho-Clinical Diagnostics is recalling Vitros XT Chemistry Products Alkp Slides, distributed nationwide in the US. The recall was initiated on 5 Aug 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Code: 1053180 UDI:(10758750008346) GEN #21; Expiry Date Range 01 Jul 2023 - 01 June 2024 GEN #23; Expiry Date Range 01 Oct 2023 - 01 Sep 2024 GEN #26; Expiry Date Range 01 Aug 2023 - 01 Dec 2024 GEN #28; Expiry Date Range 01 Aug 2024 GEN #29; Expiry Date Range 01 Sep 2024 - 01 Dec 2024 GEN #30; Expiry Date Range 02 Jan 2025
Quantity recalled402,966 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2624-2023
FDA event ID
92928
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
5 Aug 2023
FDA classified
21 Sep 2023
Reported
27 Sep 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

5 Aug 2023Recall initiated by company
21 Sep 2023Classified by FDA+47 days
27 Sep 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.

Product as listed by the FDA

VITROS XT Chemistry Products ALKP Slides

Where it was distributed

Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom

Nationwide

Questions about this recall

Why was it recalled?

Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.

Which lots are affected?

Product Code: 1053180 UDI:(10758750008346) GEN #21; Expiry Date Range 01 Jul 2023 - 01 June 2024 GEN #23; Expiry Date Range 01 Oct 2023 - 01 Sep 2024 GEN #26; Expiry Date Range 01 Aug 2023 - 01 Dec 2024 GEN #28; Expiry Date Range 01 Aug 2024 GEN #29; Expiry Date Range 01 Sep 2024 - 01 Dec 2024 GEN #30; Expiry Date Range 02 Jan 2025

Where was it sold?

Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 92928
ReportedClassRecallCompanyReasonWhere
CLASS IIVitros XT Chemistry Products Tbil-alkp Slides recalledOrtho-Clinical DiagnosticsOtherNationwide

More recalls from Ortho-Clinical Diagnostics

All 311

Other Other Medical Devices recalls

All 13,778