CLASS II ONGOING Device recall · Reported 27 Sep 2023 · FDA Enforcement Report
Vitros XT Chemistry Products Alkp Slides recalled
Ortho-Clinical Diagnostics is recalling Vitros XT Chemistry Products Alkp Slides, distributed nationwide in the US. The recall was initiated on 5 Aug 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2624-2023
- FDA event ID
- 92928
- Recalling firm
- Ortho-Clinical Diagnostics, Rochester, NY
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 5 Aug 2023
- FDA classified
- 21 Sep 2023
- Reported
- 27 Sep 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 5 Aug 2023 | Recall initiated by company | |
| 21 Sep 2023 | Classified by FDA | +47 days |
| 27 Sep 2023 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.
Product as listed by the FDA
VITROS XT Chemistry Products ALKP Slides
Where it was distributed
Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom
Questions about this recall
Why was it recalled?
Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.
Which lots are affected?
Product Code: 1053180 UDI:(10758750008346) GEN #21; Expiry Date Range 01 Jul 2023 - 01 June 2024 GEN #23; Expiry Date Range 01 Oct 2023 - 01 Sep 2024 GEN #26; Expiry Date Range 01 Aug 2023 - 01 Dec 2024 GEN #28; Expiry Date Range 01 Aug 2024 GEN #29; Expiry Date Range 01 Sep 2024 - 01 Dec 2024 GEN #30; Expiry Date Range 02 Jan 2025
Where was it sold?
Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 92928| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Vitros XT Chemistry Products Tbil-alkp Slides recalled | Ortho-Clinical Diagnostics | Other | Nationwide |
More recalls from Ortho-Clinical Diagnostics
All 311| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Chemistry Products LAC Slides recalled by Ortho-Clinical Diagnostics | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products LAC Slides recalled | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products Ca Slides recalled | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products CHE Slides (Japan) recalled | Ortho-Clinical Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Vitros Chemistry Products CHE Slides recalled | Ortho-Clinical Diagnostics | Labeling Errors | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |