CLASS II ONGOING Device recall · Reported 13 Mar 2024 · FDA Enforcement Report

Vitros XT 3400 Chemistry System recalled by Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics is recalling Vitros XT 3400 Chemistry System, distributed nationwide in the US. The recall was initiated on 26 Jan 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVITROS System Software Versions 3.8.0 or 3.8.1 UDI-DI 10758750031986 Serial Numbers: US: J34500140 J34500180 J34500257 J34500289 J34500319 J34500386 J34500422 J34500447 J34500473 J34500474 J34500510 J34500539 J34500547 J34500579 J34500588 J34500622 J34500656 J34500777 J34500818 J34500861 J34500887 J34500893 J34500896 J34500911 J34500931 OUS: J34500123 J34500128 J34500129 J34500130 J34500190 J34500261 J34500269 J34500271 J34500292 J34500301 J34500337 J34500356 J34500363 J34500515 J34500520 J34500536 J34500582 J34500609 J34500628 J34500658 J34500711 J34500793 J34500802 J34500940 J34500960 J34500962
Quantity recalled25 US; 26 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1232-2024
FDA event ID
93927
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
26 Jan 2024
FDA classified
1 Mar 2024
Reported
13 Mar 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

26 Jan 2024Recall initiated by company
1 Mar 2024Classified by FDA+35 days
13 Mar 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Nationwide

Questions about this recall

Why was it recalled?

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Which lots are affected?

VITROS System Software Versions 3.8.0 or 3.8.1 UDI-DI 10758750031986 Serial Numbers: US: J34500140 J34500180 J34500257 J34500289 J34500319 J34500386 J34500422 J34500447 J34500473 J34500474 J34500510 J34500539 J34500547 J34500579 J34500588 J34500622 J34500656 J34500777 J34500818 J34500861 J34500887 J34500893 J34500896 J34500911 J34500931 OUS: J34500123 J34500128 J34500129 J34500130 J34500190 J34500261 J34500269 J34500271 J34500292 J34500301 J34500337 J34500356 J34500363 J34500515 J34500520 J34500536 J34500582 J34500609 J34500628 J34500658 J34500711 J34500793 J34500802 J34500940 J34500960 J34500962

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 93927

This product is part of a recall event; the main page for the event is Vitros XT 7600 Integrated System recalled by Ortho-Clinical Diagnostics.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Other Medical Devices recalls

All 13,884