CLASS II Device recall · Reported 25 Sep 2013 · FDA Enforcement Report

3M Scotchcast, Wet or Dry Cast Padding recalled

3M Company - Health Care Business is recalling 3M Scotchcast, Wet or Dry Cast Padding, distributed in 33 states. The recall was initiated on 5 Aug 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog # - Lot # WDP2 - 792920, 793090, 793640 & 796078. WDP3 - 792921, 793098, 793638, 796077 & 798487. WDP4 - 792922, 793116, 793637, 796079 & 798488. WDP6 792923, 793636, 798489, 799235 & 799236
Quantity recalled1207 cases (4 puches per case, 20 rolls per pouch)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2242-2013
FDA event ID
66052
Recalling firm
3M Company - Health Care Business, Saint Paul, MN
Classification
Class II
Status
Terminated
Recall initiated
5 Aug 2013
FDA classified
18 Sep 2013
Reported
25 Sep 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

5 Aug 2013Recall initiated by company
18 Sep 2013Classified by FDA+44 days
25 Sep 2013Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

3M is conducting a voluntary product recall of selected lots of 3M" Scotchcast" Wet or Dry Cast Padding. Upon investigation of a health care provider feedback, and confirmed with internal testing, it was found that the cast padding material does not consistently repel water. Lack of repellency may result in prolonged cast dry times after exposure to wet conditions. There have been no repo

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

3M Scotchcast, Wet or Dry Cast Padding, Catalog Numbers WDP2, WDP3, WDP4, and WDP6. Intended for use in constructing casts for either wet or dry immobilization.

Where it was distributed

Worldwide Distribution - USA including AL, AZ, AR, CA, CO,CT, FL, GA, HI, IL, IN,KS,KY, LA, MD, MA, MI, MN, MO, NE, NV, NJ, NY, NC, OH, OR,PA, TN, TX, UT,VA, WA, and WI. Internationally to GERMANY, CANADA, BELGIUM, NETHERLANDS, HONG KONG, NEW ZEALAND, AUSTRALIA, INDIA, KOREA, SINGAPORE AND UNITED ARAB EMIRATES, JAPAN, JAMAICA..

AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriNorth Carolina

Questions about this recall

Why was it recalled?

3M is conducting a voluntary product recall of selected lots of 3M" Scotchcast" Wet or Dry Cast Padding. Upon investigation of a health care provider feedback, and confirmed with internal testing, it was found that the cast padding material does not consistently repel water. Lack of repellency may result in prolonged cast dry times after exposure to wet conditions. There have been no repo

Which lots are affected?

Catalog # - Lot # WDP2 - 792920, 793090, 793640 & 796078. WDP3 - 792921, 793098, 793638, 796077 & 798487. WDP4 - 792922, 793116, 793637, 796079 & 798488. WDP6 792923, 793636, 798489, 799235 & 799236

Where was it sold?

Worldwide Distribution - USA including AL, AZ, AR, CA, CO,CT, FL, GA, HI, IL, IN,KS,KY, LA, MD, MA, MI, MN, MO, NE, NV, NJ, NY, NC, OH, OR,PA, TN, TX, UT,VA, WA, and WI. Internationally to GERMANY, CANADA, BELGIUM, NETHERLANDS, HONG KONG, NEW ZEALAND, AUSTRALIA, INDIA, KOREA, SINGAPORE AND UNITED ARAB EMIRATES, JAPAN, JAMAICA..

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from 3M Company - Health Care Business

All 40

Other Other Medical Devices recalls

All 13,778