CLASS II ONGOING Device recall · Reported 10 Apr 2024 · FDA Enforcement Report
3M Tube Securement Device, Small/PO x/PO recalled
3M Company - Health Care Business is recalling 3M Tube Securement Device, Small/PO x/PO, distributed nationwide in the US. The recall was initiated on 20 Feb 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1475-2024
- FDA event ID
- 94039
- Recalling firm
- 3M Company - Health Care Business, Saint Paul, MN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 20 Feb 2024
- FDA classified
- 3 Apr 2024
- Reported
- 10 Apr 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Hip & Knee Implants
Timeline
| 20 Feb 2024 | Recall initiated by company | |
| 3 Apr 2024 | Classified by FDA | +43 days |
| 10 Apr 2024 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Manufacturing nonconformities are in some lots of the Tube Securement Device, such as the top film layer may be partially or completely missing and some devices may exhibit the appearance of an incorrect shape due to excess material not being removed from the product liner.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
3M Tube Securement Device, Small, 1.7 IN/PO x 2.4 IN/PO, REF 1500U, Sterile; and 3M Tube Securement Device, Large, 2.2 IN/PO x 3.5 IN/PO, REF 1501U, Sterile. Product is packaged 1 device per pouch, 25 pouches per primary carton, and 4 primary cartons per shipper box.
Where it was distributed
Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MN, NC, NH, NY, OH, PA, TX, VA, and WA, including PR. The countries of Australia, Austria, Canada, China, Finland, Germany, Ireland, New Zealand, Norway, Poland, Sweden, and the United Kingdom.
Nationwide CaliforniaColoradoFloridaIowaIllinoisIndianaKansasMarylandMinnesotaNorth CarolinaNew HampshireNew YorkOhioPennsylvaniaPuerto RicoTexasVirginiaWashington
Questions about this recall
Why was it recalled?
Manufacturing nonconformities are in some lots of the Tube Securement Device, such as the top film layer may be partially or completely missing and some devices may exhibit the appearance of an incorrect shape due to excess material not being removed from the product liner.
Which lots are affected?
REF 1500U, SKU 70-2011-8908-4, UDI-DI 30707387792405, Lot numbers: 33TNNN: Exp. 2025-11-09; 33TNMH: Exp. 2025-11-08; 33KNWC: Exp. 2024-06-02; 33KNJD: Exp. 2024-06-01; 33KND8: Exp. 2024-05-31. REF 1501U, SKU 70-2011-8924-1, UDI-DI 30707387792399, Lot numbers: 33TNH3: Exp. 2025-11-08; 33TNDC: Exp. 2025-11-07; 33KTDY: Exp. 2024-06-11; 33KWM3: Exp. 2024-06-15; 33KTWT: Exp. 2024-06-13; 33KWHE: Exp. 2024-06-14; 33KTLA: Exp. 2024-06-12; 33KTFP: Exp. 2024-06-11; 33KT9M: Exp. 2024-06-10; 33KP6M: Exp. 2024-06-03.
Where was it sold?
Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MN, NC, NH, NY, OH, PA, TX, VA, and WA, including PR. The countries of Australia, Austria, Canada, China, Finland, Germany, Ireland, New Zealand, Norway, Poland, Sweden, and the United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from 3M Company - Health Care Business
All 40Other Hip & Knee Implants recalls
All 3,111| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Intellijoint Navigation System: Workstation G2 Configured recalled | Intellijoint Surgical | Device Malfunction | Nationwide |
| Device | CLASS II | Trinity Non-Surgical Isolation Gown, AAMI PB70 Level 2, X-Large recalled | New York Embroidery Studio | Packaging Defects | Nationwide |
| Device | CLASS II | Axle Pin Assy Large Elbow Cobalt Chrome recalled by Howmedica Osteonics | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Triathlon Hinge Insert Size 4 5612-p-411 Triathlon Hinge recalled | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Scorpio Recessed Metal Backed Patella 3045 recalled | Howmedica Osteonics | Foreign Material | Nationwide |