CLASS II Device recall · Reported 4 May 2016 · FDA Enforcement Report

3M Steri-Vac Gas Sterilizer/Aerator recalled by 3M Company - Health Care

3M Company - Health Care Business is recalling 3M Steri-Vac Gas Sterilizer/Aerator, distributed in 15 states. The recall was initiated on 22 Jan 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel 5XL, Serial Numbers: 722612 - 722683, 820352 - 820360 Model 8XL, Serial Numbers: 351303 - 351399, 450144 - 450147
Quantity recalled74 valves

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1563-2016
FDA event ID
73117
Recalling firm
3M Company - Health Care Business, Saint Paul, MN
Classification
Class II
Status
Terminated
Recall initiated
22 Jan 2016
FDA classified
24 Apr 2016
Reported
4 May 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

22 Jan 2016Recall initiated by company
24 Apr 2016Classified by FDA+93 days
4 May 2016Published in enforcement report+10 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.

Product as listed by the FDA

3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.

Where it was distributed

AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include: Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.

AlabamaColoradoConnecticutFloridaGeorgiaIndianaNorth CarolinaNew JerseyNew YorkOklahomaOregonPennsylvaniaRhode IslandVirginiaWisconsin

Questions about this recall

Why was it recalled?

3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.

Which lots are affected?

Model 5XL, Serial Numbers: 722612 - 722683, 820352 - 820360 Model 8XL, Serial Numbers: 351303 - 351399, 450144 - 450147

Where was it sold?

AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include: Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from 3M Company - Health Care Business

All 40

Other Other Medical Devices recalls

All 13,778