CLASS II Device recall · Reported 17 Apr 2019 · FDA Enforcement Report
Abbott Architect c4000 Processing Module, List No. 02P24 recalled
Abbott Laboratories is recalling Abbott Architect c4000 Processing Module, List No. 02P24, distributed nationwide in the US. The recall was initiated on 17 Jul 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1102-2019
- FDA event ID
- 82104
- Recalling firm
- Abbott Laboratories, Irving, TX
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 17 Jul 2018
- FDA classified
- 8 Apr 2019
- Reported
- 17 Apr 2019
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Patient Monitors
Timeline
| 17 Jul 2018 | Recall initiated by company | |
| 8 Apr 2019 | Classified by FDA | +265 days |
| 17 Apr 2019 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
Product as listed by the FDA
Abbott ARCHITECT c4000 Processing Module, List No. 02P24 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitoring system that helps identify errors in aspiration.
Where it was distributed
Worldwide distribution - US Nationwide and countries Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina FASO, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Democratic Rep. of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Mauritania, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, St. Lucia, St. Vincent, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Venezuela, Vietnam, Yemen, Zambia, and Zimbabwe.
Questions about this recall
Is Abbott Architect c4000 Processing Module, List No. 02P24 recalled?
Yes. Abbott Laboratories recalled Abbott Architect c4000 Processing Module, List No. 02P24 on 17 Jul 2018. The recall is Class II and its status is Terminated. Recall number Z-1102-2019.
Why was it recalled?
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
Which lots are affected?
All serial numbers
Where was it sold?
Worldwide distribution - US Nationwide and countries Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina FASO, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Democratic Rep. of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Mauritania, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, St. Lucia, St. Vincent, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Venezuela, Vietnam, Yemen, Zambia, and Zimbabwe.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Abbott Laboratories
All 153| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Alinity hq Analyzer recalled by Abbott Laboratories | Abbott Laboratories | Device Malfunction | Nationwide |
| Device | CLASS II | Cell-dyn Ruby recalled by Abbott Laboratories | Abbott Laboratories | Device Malfunction | Nationwide |
| Device | CLASS II | Cell-dyn Ruby recalled by Abbott Laboratories | Abbott Laboratories | Device Malfunction | Nationwide |
| Device | CLASS II | Cell-dyn Ruby recalled by Abbott Laboratories | Abbott Laboratories | Device Malfunction | Nationwide |
| Device | CLASS II | Cell-dyn Ruby recalled by Abbott Laboratories | Abbott Laboratories | Device Malfunction | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |