CLASS II ONGOING Device recall · Reported 6 May 2020 · FDA Enforcement Report
ABL800 FLEX recalled by Radiometer Medical ApS
Radiometer Medical ApS is recalling ABL800 FLEX, distributed in California. The recall was initiated on 6 Mar 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1836-2020
- FDA event ID
- 85166
- Recalling firm
- Radiometer Medical ApS, Bronshoj, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 6 Mar 2020
- FDA classified
- 30 Apr 2020
- Reported
- 6 May 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 6 Mar 2020 | Recall initiated by company | |
| 30 Apr 2020 | Classified by FDA | +55 days |
| 6 May 2020 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.
Product as listed by the FDA
ABL800 FLEX model number 393-800 and 393-801, UDI:(01)05700693938004
Where it was distributed
US: in the state of California
Questions about this recall
Why was it recalled?
Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.
Which lots are affected?
Serial Numbers: 902-754R2692N002 and 902-754R2692N001
Where was it sold?
US: in the state of California
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Radiometer Medical ApS
All 24| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Radiometer ABL90 Series - Radiometer ABL90 FLEX recalled | Radiometer Medical ApS | Device Malfunction | 49 states |
| Device | CLASS II | Aqure REF 933-599 Software Versions 2.5.2 recalled | Radiometer Medical ApS | Packaging Defects | Nationwide |
| Device | CLASS II | ABL800 Flex Analyzer recalled by Radiometer Medical ApS | Radiometer Medical ApS | Other | Nationwide |
| Device | CLASS II | Radiometer, REF:956-552, PIC050- Arterial Blood Sampler 1 X 2 mL recalled | Radiometer Medical ApS | Sterility | Nationwide |
| Device | CLASS II | Radiometer ABL90 Flex Plus Analyzer with E3800 PC units recalled | Radiometer Medical ApS | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |