CLASS II Device recall · Reported 3 Jun 2020 · FDA Enforcement Report

ABL90 FLEX recalled by Radiometer Medical ApS

Radiometer Medical ApS is recalling ABL90 FLEX, distributed nationwide in the US. The recall was initiated on 21 Nov 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesABL90 FLEX Model no. 393-090: R0659 - R0977 and Spare Part Hemolyzer Units (902-980): R0526 - R0941
Quantity recalled1233 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2139-2020
FDA event ID
85390
Recalling firm
Radiometer Medical ApS, Bronshoj, N/A
Classification
Class II
Status
Terminated
Recall initiated
21 Nov 2019
FDA classified
28 May 2020
Reported
3 Jun 2020
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Other Reasons

Timeline

21 Nov 2019Recall initiated by company
28 May 2020Classified by FDA+189 days
3 Jun 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

Product as listed by the FDA

ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Where it was distributed

Worldwide distribution - US Nationwide. Also distributed OUS.

Nationwide

Questions about this recall

Why was it recalled?

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

Which lots are affected?

ABL90 FLEX Model no. 393-090: R0659 - R0977 and Spare Part Hemolyzer Units (902-980): R0526 - R0941

Where was it sold?

Worldwide distribution - US Nationwide. Also distributed OUS.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 85390
ReportedClassRecallCompanyReasonWhere
CLASS IIAbl90 Flex Plus recalled by Radiometer Medical ApSRadiometer Medical ApSOtherNationwide

More recalls from Radiometer Medical ApS

All 24

Other Other Medical Devices recalls

All 13,908