CLASS II Device recall · Reported 24 Oct 2012 · FDA Enforcement Report
AutoMate 1250 and AutoMate 2550, Part Numbers ODL25125 and ODL25255 recalled
Beckman Coulter is recalling AutoMate 1250 and AutoMate 2550, Part Numbers ODL25125 and ODL25255, distributed nationwide in the US. The recall was initiated on 11 Apr 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0056-2013
- FDA event ID
- 63155
- Recalling firm
- Beckman Coulter, Brea, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 11 Apr 2012
- FDA classified
- 12 Oct 2012
- Reported
- 24 Oct 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Masks, Gloves & PPE
Timeline
| 11 Apr 2012 | Recall initiated by company | |
| 12 Oct 2012 | Classified by FDA | +184 days |
| 24 Oct 2012 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The recall was initiated because Beckman Coulter has confirmed customer reports of secondary tubes being labeled with two different barcode labels on the AutoMate 2500 Family (previously referred to as OLA2500 Systems) units which can cause misidentification of a sample when software versions lower than version 4.1 is used.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
AutoMate 1250 and AutoMate 2550, Part Numbers ODL25125 and ODL25255. The AutoMate 1200/2500 System Series is a semi-open, pre- and post-analytical sample processing and sorting system. The base system automates the sample sorting, decapping, and archiving processes. Handling and sorting of samples includes automatic detection of the tube type and cap type (color). Optional features include an Aliquot Module for creation of labeled secondary tubes (including the detection of the presence of adequate volume for the requested aliquots) and a Recapper Module to re-seal previously decapped tubes prior to archiving.
Where it was distributed
Worldwide Distribution - USA (nationwide) and the following countries: Argentina Australia Austria Belarus Belgium China Croatia Czech Republic Denmark Finland France Germany Hungary India Ireland Israel Italy Japan Korea, Republic of Kuwait Macao Malaysia Netherlands New Zealand Norway Panama Russian Federation Slovakia Slovenia South Africa Spain Sweden Switzerland Turkey United Kingdom
Questions about this recall
Why was it recalled?
The recall was initiated because Beckman Coulter has confirmed customer reports of secondary tubes being labeled with two different barcode labels on the AutoMate 2500 Family (previously referred to as OLA2500 Systems) units which can cause misidentification of a sample when software versions lower than version 4.1 is used.
Which lots are affected?
All Serial Numbers
Where was it sold?
Worldwide Distribution - USA (nationwide) and the following countries: Argentina Australia Austria Belarus Belgium China Croatia Czech Republic Denmark Finland France Germany Hungary India Ireland Israel Italy Japan Korea, Republic of Kuwait Macao Malaysia Netherlands New Zealand Norway Panama Russian Federation Slovakia Slovenia South Africa Spain Sweden Switzerland Turkey United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter
All 334| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US) recalled | Beckman Coulter | Other | Nationwide |
| Device | CLASS II | REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) recalled | Beckman Coulter | Other | Nationwide |
| Device | CLASS II | Access Total T4 Calibrator recalled by Beckman Coulter | Beckman Coulter | Other | Nationwide |
Other Masks, Gloves & PPE recalls
All 447| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | P28 Phanton Intramedullary Nail System recalled by Paragon 28 | Paragon 28 | Sterility | Nationwide |
| Device | CLASS II | Cyberknife Treatment Deliver System equipped recalled | Accuray | Device Malfunction | Nationwide |
| Device | CLASS II | Adapter for Head Positioning recalled by Baxter Healthcare | Baxter Healthcare | Other | Nationwide |
| Device | CLASS II | Cyberknife Treatment Deliver System equipped recalled | Accuray | Device Malfunction | Nationwide |
| Device | CLASS II | Halyard Purezero Ultraviolet Nitrile Exam Gloves Reference recalled | O&M Halyard | Other | Nationwide |