CLASS II ONGOING Device recall · Reported 22 Mar 2023 · FDA Enforcement Report

Bard Mission Disposable Core Biopsy Instrument Kit recalled

Bard Peripheral Vascular is recalling Bard Mission Disposable Core Biopsy Instrument Kit, distributed nationwide in the US. The recall was initiated on 28 Feb 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI/Lot(Expiration Date): 1610MSK/00801741142567/1445630(11/28/2024), 1449980(12/28/2024), 1451134(1/28/2025); 1816MSK/00801741097065/1447510(12/28/2024), 1447511(12/28/2024), 1447512(12/28/2024), 1447513(12/28/2024), 1448118(12/28/2024), 1448119(12/28/2024), 1448120(12/28/2024), 1448121(12/28/2024), 1448743(12/28/2024), 1450413(12/28/2024), 1450414(12/28/2024), 1451075(1/28/2025), 1451076(1/28/2025), 1451621(1/28/2025), 1451622(1/28/2025), 1451625(1/28/2025), 1451626(1/28/2025), 1453516(1/28/2025), 1453517(1/28/2025), 1454295(1/28/2025), 1456850(2/28/2025), 1456856(2/28/2025), 1457030(2/28/2025)
Quantity recalled38,635

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1260-2023
FDA event ID
91638
Recalling firm
Bard Peripheral Vascular, Tempe, AZ
Classification
Class II
Status
Ongoing
Recall initiated
28 Feb 2023
FDA classified
16 Mar 2023
Reported
22 Mar 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

28 Feb 2023Recall initiated by company
16 Mar 2023Classified by FDA+16 days
22 Mar 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK

Where it was distributed

Worldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO, PA, TX, NC, CA, ME, MO, NV, MI, IL, NY, VA, AZ, NH, LA, WI, KY, MA, WA, CT, SD, OK, DC, IA, AR, GA, NE, NM, IN, NJ, MN, PR, AL, AK, SC, KS, OR, ID, VT, MT, WV, ND, TN, MS, HI, WY, RI and OUS countries of: Canada, Korea, Thailand and Taiwan. OUS: AR, AU, BE, CA, CN, CO, HK, IN, JP, KR, TH, TW

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Why was it recalled?

The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.

Which lots are affected?

REF/UDI-DI/Lot(Expiration Date): 1610MSK/00801741142567/1445630(11/28/2024), 1449980(12/28/2024), 1451134(1/28/2025); 1816MSK/00801741097065/1447510(12/28/2024), 1447511(12/28/2024), 1447512(12/28/2024), 1447513(12/28/2024), 1448118(12/28/2024), 1448119(12/28/2024), 1448120(12/28/2024), 1448121(12/28/2024), 1448743(12/28/2024), 1450413(12/28/2024), 1450414(12/28/2024), 1451075(1/28/2025), 1451076(1/28/2025), 1451621(1/28/2025), 1451622(1/28/2025), 1451625(1/28/2025), 1451626(1/28/2025), 1453516(1/28/2025), 1453517(1/28/2025), 1454295(1/28/2025), 1456850(2/28/2025), 1456856(2/28/2025), 1457030(2/28/2025)

Where was it sold?

Worldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO, PA, TX, NC, CA, ME, MO, NV, MI, IL, NY, VA, AZ, NH, LA, WI, KY, MA, WA, CT, SD, OK, DC, IA, AR, GA, NE, NM, IN, NJ, MN, PR, AL, AK, SC, KS, OR, ID, VT, MT, WV, ND, TN, MS, HI, WY, RI and OUS countries of: Canada, Korea, Thailand and Taiwan. OUS: AR, AU, BE, CA, CN, CO, HK, IN, JP, KR, TH, TW

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Bard Peripheral Vascular

All 139

Other Other Medical Devices recalls

All 13,908