CLASS II Device recall · Reported 29 May 2013 · FDA Enforcement Report

Brilliance CT 16-Slice Water, System recalled by Philips Medical Systems

Philips Medical Systems (Cleveland) is recalling Brilliance CT 16-Slice Water, System, distributed nationwide in the US. The recall was initiated on 12 Mar 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSystem Code #728255, Serial #: 9702, 9703, 9704, 9705, 9706, 9707, 9709, 9710, 9713, 9714, 9717, 9718, 9719, 9721, 9722, 9723, 9724, 9725, 9727, 9730, 9731, 9732, 9733, 9734, 9735, 9738, 9739, 9740, 9741, 9742, 9743, 9746, 9747, 9749, 9752, 9753, 9756, 9758, 9759, 9760, 9761, 9762, 9765, 9766, 9768, 9769, 9770, 9772, 9774, 9778, 9780, 9781, 9783, 9784, 9786, 9787, 9790, 9791, 9792, 9793, 9796, 9797, 9801, 9805, 9807, 9808 & 9810. System Code #728250, Serial #: 2556, 2557, 2560, 2562, 2563, 2565, 2567, 2568, 2569, 2570, 2571, 2574, 2576, 2577, 2579, 2580, 2581, 2584, 2585, 2587, 2588, 2590, 2591, 2592, 2593, 2594, 2595, 2597, 9738 & 9741. System Code #728245, Serial #: 2562, 2564, 2567, 2573, 2576, 2582, 2584, 2585, 2586, 2587, 2589, 2590, 2594, 2596, 2597, 3014, 3091, 3105, 3151, 3217, 3235, 3287, 3303, 3312, 3339, 3340, 3341, 3342, 3343, 3344, 3346, 3348, 3351, 3353, 3355, 3356, 3357, 3358, 3359, 3360, 3363, 3364, 3366, 3369, 3371, 3373, 3374, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3385, 3388, 3392, 3393, 3394, 3397, 3400, 3404, 3406, 3407, 3408, 3410, 3413, 3414, 3415, 3416, 3419, 3423, 3425, 3427, 3429, 3431, 3432, 3433, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3443, 3445, 3446, 3448, 3449, 3451, 3452, 3453, 3454, 3455, 3458, 3459, 3461, 3462, 3464, 3468, 3469, 3470, 3471, 3473, 3474, 3475, 3477, 3478, 3479, 3481, 3482, 3483, 3484, 3485, 3486, 3488, 3489, 3491, 3492, 3493, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3517, 9716, 9718, 9719, 9724, 9732, 9741, 9743, 9746, 9749, 9778, 9780, 9787, 9790, 9797, 9801, 9808 & 9810. System Code #728120, Serial #: 2502, 2510, 2513, 2516, 2517, 2518, 2525, 2528, 2531, 2533, 2535, 2539, 2546, 3006, 3008, 3013, 3014, 3021, 3025, 3026, 3027, 3028, 3029, 3030, 3035, 3037, 3039, 3040, 3043, 3044, 3047, 3049, 3050, 3051, 3053, 3054, 3056, 3059, 3061, 3062, 3063, 3064, 3068, 3069, 3070, 3072, 3073, 3075, 3076, 3077, 3078, 3079, 3081, 3082, 3084, 3085, 3087, 3088, 3089, 3091, 3093, 3094, 3095, 3097, 3099, 3100, 3103, 3104, 3105, 3106, 3107, 3108, 3110, 3111, 3112, 3113, 3114, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3124, 3125, 3126, 3127, 3129, 3130, 3131, 3132, 3135, 3136, 3137, 3138, 3141, 3143, 3144, 3146, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3171, 3172, 3174, 3176, 3177, 3178, 3179, 3180, 3182, 3183, 3185, 3187, 3188, 3189, 3190, 3192, 3194, 3195, 3196, 3198, 3199, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3211, 3212, 3213, 3214, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3229, 3230, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3245, 3246, 3247, 3248, 3250, 3251, 3252, 3253, 3254, 3255, 3258, 3261, 3262, 3264, 3265, 3266, 3268, 3269, 3270, 3271, 3273, 3274, 3277, 3279, 3280, 3281, 3283, 3284, 3285, 3287, 3289, 3290, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3302, 3303, 3304, 3305, 3307, 3308, 3310, 3311, 3312, 3315, 3316, 3318, 3319, 3320, 3321, 3322, 3324, 3328, 3329, 3330, 3367 & 9760. System Code #728125, Serial #: 2501, 2505, 2508, 2509, 2513, 2515, 2516, 2522, 2523, 2524, 2525, 2526, 2529, 2530, 2531, 2533, 2534, 2535, 2539, 2540, 2542, 2543, 2548 & 2549.
Quantity recalled515 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1338-2013
FDA event ID
64713
Recalling firm
Philips Medical Systems (Cleveland), Cleveland, OH
Classification
Class II
Status
Terminated
Recall initiated
12 Mar 2013
FDA classified
20 May 2013
Reported
29 May 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

12 Mar 2013Recall initiated by company
20 May 2013Classified by FDA+69 days
29 May 2013Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information.

Product as listed by the FDA

Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc., 595 Miner Road, Cleveland, OH 44143 Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Where it was distributed

Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, WV and WY; countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Denmark, France, Germany, Gibraltar, India, Iraq, Ireland, Israel, Italy, Japan, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichigan

Questions about this recall

Why was it recalled?

Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information.

Which lots are affected?

System Code #728255, Serial #: 9702, 9703, 9704, 9705, 9706, 9707, 9709, 9710, 9713, 9714, 9717, 9718, 9719, 9721, 9722, 9723, 9724, 9725, 9727, 9730, 9731, 9732, 9733, 9734, 9735, 9738, 9739, 9740, 9741, 9742, 9743, 9746, 9747, 9749, 9752, 9753, 9756, 9758, 9759, 9760, 9761, 9762, 9765, 9766, 9768, 9769, 9770, 9772, 9774, 9778, 9780, 9781, 9783, 9784, 9786, 9787, 9790, 9791, 9792, 9793, 9796, 9797, 9801, 9805, 9807, 9808 & 9810. System Code #728250, Serial #: 2556, 2557, 2560, 2562, 2563, 2565, 2567, 2568, 2569, 2570, 2571, 2574, 2576, 2577, 2579, 2580, 2581, 2584, 2585, 2587, 2588, 2590, 2591, 2592, 2593, 2594, 2595, 2597, 9738 & 9741. System Code #728245, Serial #: 2562, 2564, 2567, 2573, 2576, 2582, 2584, 2585, 2586, 2587, 2589, 2590, 2594, 2596, 2597, 3014, 3091, 3105, 3151, 3217, 3235, 3287, 3303, 3312, 3339, 3340, 3341, 3342, 3343, 3344, 3346, 3348, 3351, 3353, 3355, 3356, 3357, 3358, 3359, 3360, 3363, 3364, 3366, 3369, 3371, 3373, 3374, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3385, 3388, 3392, 3393, 3394, 3397, 3400, 3404, 3406, 3407, 3408, 3410, 3413, 3414, 3415, 3416, 3419, 3423, 3425, 3427, 3429, 3431, 3432, 3433, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3443, 3445, 3446, 3448, 3449, 3451, 3452, 3453, 3454, 3455, 3458, 3459, 3461, 3462, 3464, 3468, 3469, 3470, 3471, 3473, 3474, 3475, 3477, 3478, 3479, 3481, 3482, 3483, 3484, 3485, 3486, 3488, 3489, 3491, 3492, 3493, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3517, 9716, 9718, 9719, 9724, 9732, 9741, 9743, 9746, 9749, 9778, 9780, 9787, 9790, 9797, 9801, 9808 & 9810. System Code #728120, Serial #: 2502, 2510, 2513, 2516, 2517, 2518, 2525, 2528, 2531, 2533, 2535, 2539, 2546, 3006, 3008, 3013, 3014, 3021, 3025, 3026, 3027, 3028, 3029, 3030, 3035, 3037, 3039, 3040, 3043, 3044, 3047, 3049, 3050, 3051, 3053, 3054, 3056, 3059, 3061, 3062, 3063, 3064, 3068, 3069, 3070, 3072, 3073, 3075, 3076, 3077, 3078, 3079, 3081, 3082, 3084, 3085, 3087, 3088, 3089, 3091, 3093, 3094, 3095, 3097, 3099, 3100, 3103, 3104, 3105, 3106, 3107, 3108, 3110, 3111, 3112, 3113, 3114, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3124, 3125, 3126, 3127, 3129, 3130, 3131, 3132, 3135, 3136, 3137, 3138, 3141, 3143, 3144, 3146, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3171, 3172, 3174, 3176, 3177, 3178, 3179, 3180, 3182, 3183, 3185, 3187, 3188, 3189, 3190, 3192, 3194, 3195, 3196, 3198, 3199, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3211, 3212, 3213, 3214, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3229, 3230, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3245, 3246, 3247, 3248, 3250, 3251, 3252, 3253, 3254, 3255, 3258, 3261, 3262, 3264, 3265, 3266, 3268, 3269, 3270, 3271, 3273, 3274, 3277, 3279, 3280, 3281, 3283, 3284, 3285, 3287, 3289, 3290, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3302, 3303, 3304, 3305, 3307, 3308, 3310, 3311, 3312, 3315, 3316, 3318, 3319, 3320, 3321, 3322, 3324, 3328, 3329, 3330, 3367 & 9760. System Code #728125, Serial #: 2501, 2505, 2508, 2509, 2513, 2515, 2516, 2522, 2523, 2524, 2525, 2526, 2529, 2530, 2531, 2533, 2534, 2535, 2539, 2540, 2542, 2543, 2548 & 2549.

Where was it sold?

Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, WV and WY; countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Denmark, France, Germany, Gibraltar, India, Iraq, Ireland, Israel, Italy, Japan, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips Medical Systems (Cleveland)

All 315

Other Imaging Equipment recalls

All 2,336