CLASS II Device recall · Reported 6 Nov 2013 · FDA Enforcement Report
Carestream DRX-Revolution Mobile X-ray System recalled
Carestream Health is recalling Carestream DRX-Revolution Mobile X-ray System, distributed nationwide in the US. The recall was initiated on 6 Jun 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0126-2014
- FDA event ID
- 65469
- Recalling firm
- Carestream Health, Rochester, NY
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 Jun 2013
- FDA classified
- 31 Oct 2013
- Reported
- 6 Nov 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Imaging Equipment
Timeline
| 6 Jun 2013 | Recall initiated by company | |
| 31 Oct 2013 | Classified by FDA | +147 days |
| 6 Nov 2013 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.
Product as listed by the FDA
Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.
Where it was distributed
Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN.
Nationwide AlabamaArizonaCaliforniaColoradoFloridaGeorgiaIowaIllinoisIndianaKansasLouisianaMassachusettsMichiganMinnesotaNorth CarolinaNew JerseyNew YorkOhioOklahomaPennsylvania
Questions about this recall
Is Carestream DRX-Revolution Mobile X-ray System recalled?
Yes. Carestream Health recalled Carestream DRX-Revolution Mobile X-ray System on 6 Jun 2013. The recall is Class II and its status is Terminated. Recall number Z-0126-2014.
Why was it recalled?
Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.
Which lots are affected?
Serial Numbers: 111, 114, 119-123, 125-126, 134, 142-143, 151-155, 161-169, 173-178,180-183, 188-192, 196-209, 211, 221-232, 235-249, 254, 261-270, 275-276, 279-286, 305, 312, 316-326, 330-337, 339-349, 353, 355, 357-365, 372-373, 375-376, 379-385, 387-388, 394-401, 403-405, 409-410, 412-415, 417, 420-421, 423, 431-435, 441-446, 451-461, 464-470, and 482-490
Where was it sold?
Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Carestream Health
All 86| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DRX Revolution Mobile X-Ray System recalled by Carestream Health | Carestream Health | Device Malfunction | Nationwide |
| Device | CLASS II | The DRX-Revolution Mobile X-Ray System is a diagnostic recalled | Carestream Health | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Drx-Compass/Dr-fit X-ray Systems with Firmware recalled | Carestream Health | Other | Nationwide |
| Device | CLASS II | Kodak DirectView DR3000/3500 recalled by Carestream Health | Carestream Health | Lead & Heavy Metals | IA |
| Device | CLASS II | OnSight 3D Extremity System- X-Ray, Tomography Computed System recalled | Carestream Health | Other | Nationwide |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device is recalled | Canon Medical System, USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederland B.v | Lead & Heavy Metals |