CLASS II ONGOING Device recall · Reported 13 Nov 2024 · FDA Enforcement Report

Cartesion Prime, Digital Pet-ct recalled by Canon Medical System, USA

Canon Medical System USA is recalling Cartesion Prime, Digital Pet-ct, distributed in 18 states. The recall was initiated on 17 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029
Quantity recalled27

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0284-2025
FDA event ID
95471
Recalling firm
Canon Medical System USA, Tustin, CA
Classification
Class II
Status
Ongoing
Recall initiated
17 Sep 2024
FDA classified
1 Nov 2024
Reported
13 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Sep 2024Recall initiated by company
1 Nov 2024Classified by FDA+45 days
13 Nov 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.

Product as listed by the FDA

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Where it was distributed

US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA

AlabamaArkansasArizonaCaliforniaColoradoIowaIllinoisIndianaKansasLouisianaMarylandMinnesotaMississippiNorth CarolinaNevadaTennesseeVirginiaWisconsin

Questions about this recall

Why was it recalled?

When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.

Which lots are affected?

UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029

Where was it sold?

US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Canon Medical System USA

All 50

Other Imaging Equipment recalls

All 2,336