CLASS II ONGOING Device recall · Reported 6 Sep 2023 · FDA Enforcement Report
The Chevalier Valvulotome is a device used to disrupt recalled
LeMaitre Vascular is recalling The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be. The recall was initiated on 13 Jul 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2468-2023
- FDA event ID
- 92825
- Recalling firm
- LeMaitre Vascular, Burlington, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 Jul 2023
- FDA classified
- 28 Aug 2023
- Reported
- 6 Sep 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Foreign Material
- Device type
- Other Medical Devices
Timeline
| 13 Jul 2023 | Recall initiated by company | |
| 28 Aug 2023 | Classified by FDA | +46 days |
| 6 Sep 2023 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Burrs (molding flash) on the cutter may tear the harvested vien
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed bypass procedure. The Chevalier Valvulotome consists of a 70 cm stainless steel shaft with two tips: 1) the cutting tip is shaped like an inverted tulip with four teeth at its base. This tip is designed to cut the vein valve cusps. 2) the other tip is atraumatic and used to test the patency and the course of the vein before the valvulotomy. The Chevalier Valvulotome is available in three tip diameters: 2.5 mm (Catalog # 40025); 3.0 mm (Catalog # 40030); and 3.5 mm (Catalog # 40035)
Where it was distributed
International distribution to the countries of Belgium, France, Germany, and United Kingdom.
Questions about this recall
Why was it recalled?
Burrs (molding flash) on the cutter may tear the harvested vien
Which lots are affected?
(1)Catalog # 40025 UDI-DI: 00840663110469; (2)Catalog # 40030 UDI-DI: 00840663110476; (3) Catalog # 40035 UDI-DI: 00840663110483 Lot Numbers: All lots starting with CHV. Only lots beginning with CHV are being recalled. Older lots (with CL lot prefixes) are not being recalled. REF (Catalog #) LOT Catalog Number: (1)40025 Lot Number: CHV1003, CHV1006, CHV1022 (2) 40030 Lot Number: CHV1002, CHV1004, (3) 40035 Lot Number: CHV1001, CHV1005
Where was it sold?
International distribution to the countries of Belgium, France, Germany, and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from LeMaitre Vascular
All 71| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TufTex¿ Over-the-Wire Embolectomy Catheter recalled | LeMaitre Vascular | Other | Nationwide |
| Device | CLASS II | Artegraft Vascular Graft recalled by LeMaitre Vascular | LeMaitre Vascular | Packaging Defects | Nationwide |
| Device | CLASS II | Artegraft¿ Collagen Vascular Graft recalled by LeMaitre Vascular | LeMaitre Vascular | Packaging Defects | AR |
| Device | CLASS II | Artegraft Collagen Vascular Graft recalled by LeMaitre Vascular | LeMaitre Vascular | Labeling Errors | NJ |
| Device | CLASS II | Artegraft Collagen Vascular Grafts recalled by LeMaitre Vascular | LeMaitre Vascular | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |