CLASS II ONGOING Device recall · Reported 1 Jul 2026 · FDA Enforcement Report
Convenience kits containing select SKUs of Polycarbonate recalled
Medline Industries is recalling Convenience kits containing select SKUs of Polycarbonate Colored Syringes, distributed nationwide in the US. The recall was initiated on 27 Feb 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2581-2026
- FDA event ID
- 99062
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 27 Feb 2026
- FDA classified
- 25 Jun 2026
- Reported
- 1 Jul 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Syringes & Needles
Timeline
| 27 Feb 2026 | Recall initiated by company | |
| 25 Jun 2026 | Classified by FDA | +118 days |
| 1 Jul 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Product as listed by the FDA
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAM PACK-LF DYNJ21574J JUDKINS PACK DYNJ51126
Where it was distributed
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIllinoisKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiMontanaNebraska
Questions about this recall
Why was it recalled?
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Which lots are affected?
DYNJ21574J UDI-DI 10889942153497 Lots 21ALA586 21CLA398 21ELA408 21FLA144 21FLA788 21HLA420 21ILA431 DYNJ51126 UDI-DI 10889942122820 Lots 21ADA917
Where was it sold?
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
15 other products · FDA event 99062This product is part of a recall event; the main page for the event is Convenience kits containing select SKUs of Polycarbonate recalled.
More recalls from Medline Industries
All 707| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Major Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Peds Sinus Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Mfc Craniotomy Pack recalled | Medline Industries | Other | Nationwide |
Other Syringes & Needles recalls
All 1,066| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Embosphere 500-700 micrometer Syringe Sterile Steam recalled | Merit Medical Systems | Sterility | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
| Device | CLASS II | Medrad¿ Stellant Flex Syringe Kit used recalled by Bayer Medical Care | Bayer Medical Care | Sterility | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |