CLASS II ONGOING Device recall · Reported 4 Mar 2026 · FDA Enforcement Report

Convenience kits used for dialysis maintenance Medline Add recalled

Medline Industries is recalling Convenience kits used for dialysis maintenance Medline Add A Cath Dialysis Kit, distributed nationwide in the US. The recall was initiated on 8 Jan 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesMedline ADD A CATH DIALYSIS KIT SKU ECVC8415A UDI-DI (ea) 10653160383628 UDI-DI (case) 00653160383621 Kit lots 2024121390 2025033190 Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 UDI-DI (ea) 10653160368755 UDI-DI (case) 00653160368758 Kit lot 2025041090 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A UDI-DI (ea) 10653160374978 UDI-DI (case) 00653160374971 Kit lots 2024051490 2025021290 2025051390 2025091090 Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 UDI-DI (ea) 10653160375845 UDI-DI (case) 00653160375848 Kit lots 2024060690 2024101690 2024103190 2025010890 2025040190 Medline DIALYSIS CHANGE KIT SKU EBSI1453A UDI-DI (ea) 10653160383475 UDI-DI (case) 00653160383478 Kit lots 2024050790 2024051390 2024082290 2024111990 2025013090 2025032090 2025050190 2025062790 Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 UDI-DI (ea) 10653160368922 UDI-DI (case) 00653160368925 Kit lots 2025040401 2025081401 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741 UDI-DI (ea) 10653160374527 UDI-DI (case) 00653160374520 Kit lots 2024061780 2024070180 2024081280 2024111290 2025012490 2025032690 2025051390 2025061790 2025071490 2025072190
Quantity recalled14,525

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1465-2026
FDA event ID
98333
Recalling firm
Medline Industries, Northfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
8 Jan 2026
FDA classified
24 Feb 2026
Reported
4 Mar 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

8 Jan 2026Recall initiated by company
24 Feb 2026Classified by FDA+47 days
4 Mar 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Which lots are affected?

Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A UDI-DI (ea) 10653160383628 UDI-DI (case) 00653160383621 Kit lots 2024121390 2025033190 Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 UDI-DI (ea) 10653160368755 UDI-DI (case) 00653160368758 Kit lot 2025041090 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A UDI-DI (ea) 10653160374978 UDI-DI (case) 00653160374971 Kit lots 2024051490 2025021290 2025051390 2025091090 Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 UDI-DI (ea) 10653160375845 UDI-DI (case) 00653160375848 Kit lots 2024060690 2024101690 2024103190 2025010890 2025040190 Medline DIALYSIS CHANGE KIT SKU EBSI1453A UDI-DI (ea) 10653160383475 UDI-DI (case) 00653160383478 Kit lots 2024050790 2024051390 2024082290 2024111990 2025013090 2025032090 2025050190 2025062790 Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 UDI-DI (ea) 10653160368922 UDI-DI (case) 00653160368925 Kit lots 2025040401 2025081401 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741 UDI-DI (ea) 10653160374527 UDI-DI (case) 00653160374520 Kit lots 2024061780 2024070180 2024081280 2024111290 2025012490 2025032690 2025051390 2025061790 2025071490 2025072190

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 98333

This product is part of a recall event; the main page for the event is Medline Kits containing Tego Connectors Medline Dialysis ON/OFF KIT recalled.