Medline kits containing Tego Connectors Medline Apheresis recalled

Medline Industries is recalling Medline kits containing Tego Connectors Medline Apheresis Subqport Access Kit, distributed nationwide in the US. The recall was initiated on 8 Jan 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GBQ405 SKU DYNJ63347A UDI-DI (ea) 10193489278064 UDI-DI (case) 40193489278065 Kit lots 24AMC874 24AME139 24AMH820 24BMF097 24CMF691 24GMB769 24GMK036 24IMC124 24LMA110 25AMJ634 25GMD386 25IMB510 25JMI708
Quantity recalled1212

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1466-2026
FDA event ID
98333
Recalling firm
Medline Industries, Northfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
8 Jan 2026
FDA classified
24 Feb 2026
Reported
4 Mar 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

8 Jan 2026Recall initiated by company
24 Feb 2026Classified by FDA+47 days
4 Mar 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Which lots are affected?

SKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GBQ405 SKU DYNJ63347A UDI-DI (ea) 10193489278064 UDI-DI (case) 40193489278065 Kit lots 24AMC874 24AME139 24AMH820 24BMF097 24CMF691 24GMB769 24GMK036 24IMC124 24LMA110 25AMJ634 25GMD386 25IMB510 25JMI708

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 98333

This product is part of a recall event; the main page for the event is Medline Kits containing Tego Connectors Medline Dialysis ON/OFF KIT recalled.

More recalls from Medline Industries

All 709

Other Other Medical Devices recalls

All 13,908