CLASS II ONGOING Device recall · Reported 7 Feb 2024 · FDA Enforcement Report

Cortera Spinal Fixation System recalled by XTANT Medical Holdings

Xtant Medical Holdings is recalling Cortera Spinal Fixation System, distributed in Michigan. The recall was initiated on 28 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00810076564005, Lot: AP23000143
Quantity recalled6

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0846-2024
FDA event ID
93738
Recalling firm
Xtant Medical Holdings, Belgrade, MT
Classification
Class II
Status
Ongoing
Recall initiated
28 Nov 2023
FDA classified
29 Jan 2024
Reported
7 Feb 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

28 Nov 2023Recall initiated by company
29 Jan 2024Classified by FDA+62 days
7 Feb 2024Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Product as listed by the FDA

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Where it was distributed

US distribution in MI.

Michigan

Questions about this recall

Why was it recalled?

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Which lots are affected?

UDI-DI: 00810076564005, Lot: AP23000143

Where was it sold?

US distribution in MI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.