CLASS I ONGOING Device recall · Reported 20 May 2026 · FDA Enforcement Report

Custom Convenience Kit containing Medline Namic recalled

Windstone Medical Packaging is recalling Custom Convenience Kit containing Medline Namic Angiographic Control Syringes with, distributed nationwide in the US. The recall was initiated on 2 Apr 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesPack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088
Quantity recalled150 kits (US only)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2066-2026
FDA event ID
98741
Recalling firm
Windstone Medical Packaging, Billings, MT
Classification
Class I
Status
Ongoing
Recall initiated
2 Apr 2026
FDA classified
8 May 2026
Reported
20 May 2026
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

2 Apr 2026Recall initiated by company
8 May 2026Classified by FDA+36 days
20 May 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.

Product as listed by the FDA

Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F

Where it was distributed

US Nationwide distribution in the state of Arizona.

Nationwide Arizona

Questions about this recall

Why was it recalled?

Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.

Which lots are affected?

Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088

Where was it sold?

US Nationwide distribution in the state of Arizona.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Windstone Medical Packaging

All 337

Other Syringes & Needles recalls

All 1,065