CLASS I Device recall · Reported 9 Apr 2014 · FDA Enforcement Report

Draeger PS500 Power Supply Unit of the Infinity ACS recalled

Draeger Medical Systems is recalling Draeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita, distributed nationwide in the US. The recall was initiated on 7 Feb 2014 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesall lots
Quantity recalled1544

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1298-2014
FDA event ID
67502
Recalling firm
Draeger Medical Systems, Telford, PA
Classification
Class I
Status
Terminated
Recall initiated
7 Feb 2014
FDA classified
28 Mar 2014
Reported
9 Apr 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Feb 2014Recall initiated by company
28 Mar 2014Classified by FDA+49 days
9 Apr 2014Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Draeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), Evita V500 and /or Babylog VN500 Ventilators, used for ventilation during short term transport of patients within a hospital.

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.

Which lots are affected?

all lots

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Draeger Medical Systems

All 12

Other CPAP & Ventilators recalls

All 554