CLASS II Device recall · Reported 18 Jul 2012 · FDA Enforcement Report

Drager PT 4000 Phototherapy System recalled by Draeger Medical Systems

Draeger Medical Systems is recalling Drager PT 4000 Phototherapy System, distributed nationwide in the US. The recall was initiated on 2 Jul 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.
Quantity recalled65

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1950-2012
FDA event ID
62470
Recalling firm
Draeger Medical Systems, Telford, PA
Classification
Class II
Status
Terminated
Recall initiated
2 Jul 2012
FDA classified
6 Jul 2012
Reported
18 Jul 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

2 Jul 2012Recall initiated by company
6 Jul 2012Classified by FDA+4 days
18 Jul 2012Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Drager PT 4000 Phototherapy System

Where it was distributed

Nationwide Distribution-USA (nationwide) and Puerto Rico.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b

Which lots are affected?

Catalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.

Where was it sold?

Nationwide Distribution-USA (nationwide) and Puerto Rico.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.