CLASS II Device recall · Reported 9 Nov 2016 · FDA Enforcement Report

Empowr PS Knee System Box Cut Guide recalled by Encore Medical

Encore Medical is recalling Empowr PS Knee System Box Cut Guide. The recall was initiated on 19 Sep 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes202351L01, 202434L01, 203779L01, 202351L02, 202434L02, 202351L03, 202434L03, 203779L03, 202351L04, 202434L04, 202351L05, 202434L05, 202351L06, 202434L06, 203779L06, 202351L07, 202434L07, 203779L07, 202351L08, 202434L08, 203779L08, 202351L09, 202434L09, 202351L10, 202434L10, 203203L01
Quantity recalled436 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0325-2017
FDA event ID
75214
Recalling firm
Encore Medical, Austin, TX
Classification
Class II
Status
Terminated
Recall initiated
19 Sep 2016
FDA classified
28 Oct 2016
Reported
9 Nov 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

19 Sep 2016Recall initiated by company
28 Oct 2016Classified by FDA+39 days
9 Nov 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.

Product as listed by the FDA

Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.

Where it was distributed

US

Questions about this recall

Is Empowr PS Knee System Box Cut Guide recalled?

Yes. Encore Medical recalled Empowr PS Knee System Box Cut Guide on 19 Sep 2016. The recall is Class II and its status is Terminated. Recall number Z-0325-2017.

Why was it recalled?

It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.

Which lots are affected?

202351L01, 202434L01, 203779L01, 202351L02, 202434L02, 202351L03, 202434L03, 203779L03, 202351L04, 202434L04, 202351L05, 202434L05, 202351L06, 202434L06, 203779L06, 202351L07, 202434L07, 203779L07, 202351L08, 202434L08, 203779L08, 202351L09, 202434L09, 202351L10, 202434L10, 203203L01

Where was it sold?

US

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Encore Medical

All 99

Other Hip & Knee Implants recalls

All 3,097